cs.CVOct 29, 2026

Neural Blind Deconvolution for Image Scanning Microscopy

Authors: Eliot Guez, Nathan Buskulic, Lisa Cuneo, Luca Calatroni, Giuseppe Vicidomini

Organizations: MaLGa Center, DIBRIS, University of Genoa, Italy · CSLM, Italian Institute of Technology, Italy · MMS, Italian Institute of Technology, Italy

Abstract

Image scanning microscopy (ISM) acquires one scanned image per detector element, which are then fused into a single image by multi-image deconvolution to obtain a super-resolved image. Its accuracy, however, depends on point spread function (PSF) models that sample- and system-induced aberrations often invalidate. Since differences between the scanned images encode aberration information, we propose a training-free blind Deep Image Prior (DIP) framework that jointly estimates the specimen and the detector-dependent PSFs. The PSFs are parametrised through the detector geometry and shared excitation and emission PSFs, reducing the unknowns while preserving the physical ISM model, and a Gaussian pre-initialisation stabilises the blind optimisation. On experimental tubulin data, blind DIP recovers finer filamentous structures than multi-image deconvolution and DIP with fixed theoretical PSFs, while adapting the PSFs to the measurements.

Figures & tables

Explore similar work

May 29, 2026

Self-Tuning Regularization for Image Scanning Microscopy

Image Scanning Microscopy (ISM) is a fluorescence imaging technique that combines detector-array acquisition and computational reconstruction to achieve the theoretical resolution of an ideal confocal microscope, i.e., one operating with an infinitesimally small pinhole, while maintaining high signal-to-noise ratio. Among the reconstruction methods for obtaining the super-resolved image, multi-image deconvolution (MID) and its extension aimed at preserving the optical sectioning capability of confocal microscopy, known as super-resolution sectioning ISM (s2^2ISM), are among the most widely used approaches. Both methods rely on Richardson--Lucy-type iterative schemes, whose semi-convergent behavior requires early stopping and often leads to noise amplification and reconstruction artifacts. In this work, we introduce a self-tuning explicit regularization framework for both MID and s2^2ISM reconstruction. Within a Bayesian maximum a posteriori formulation, we combine a multi-frame Poisson data fidelity term with explicit regularization, considering ℓ1\ell_1 and smoothed total variation penalties as representative examples. We further develop an automatic and ground-truth-free strategy for regularization parameter selection by adapting the residual whiteness principle to the multi-frame Poisson setting and introducing a spectral high-pass extension tailored to s2^2ISM. The resulting framework enables stable reconstructions without empirical stopping rules. To demonstrate the proposed framework, we consider first-order optimization schemes based on proximal gradient and mirror descent methods with adaptive backtracking strategies. Experiments on simulated and real fluorescence ISM datasets demonstrate improved reconstruction stability and image quality with respect to unregularized approaches, while enabling robust super-resolution and optical sectioning in low-photon conditions.
Jun 26, 2026

A Zero-Shot Deep Image Prior Framework for Denoising and Deconvolution in Fluorescence Microscopy

Fluorescence microscopy images are degraded by noise and diffraction-induced blur, which compromise structural fidelity and limit quantitative analysis. Supervised deep learning methods achieve impressive restoration performance but require large-scale paired datasets that are difficult to obtain in practice. To address this issue, we propose SDIP, a zero-shot deep image prior (DIP) framework that sequentially performs denoising and deconvolution without external training data. An aSeqDIP-based module first suppresses noise while preserving fine structures through sequential autoencoding regularization. In the deconvolution stage, a wavelet-based background correction step is incorporated before the proposed RLG-DIP module performs artifact-reduced deconvolution. RLG-DIP uses the Richardson-Lucy deconvolution result as a physically consistent guidance prior, integrating the imaging model with the implicit prior of DIP to stabilize the ill-posed deconvolution process. Experiments on the BioSR dataset across multiple cellular structures demonstrate that SDIP improves both signal-to-noise ratio and resolution, achieving superior visual quality and improved quantitative performance on most evaluated structures. The proposed framework may also provide useful insights for designing physically guided DIP methods for other inverse problems.
May 13, 2026

DeepFilters: Scattering-Aware Pupil Engineering with Learned Digital Filter Reconstruction for Extended Depth of Field Microscopy

Extended depth of field microscopy encodes axial information into a single acquisition through engineered point spread functions, but conventional and deep optics approaches are subject to degradation in scattering tissue. We introduce DeepFilters, a scattering-aware deep optics framework that jointly optimizes a parameterized pupil filter and a digital-filter-based reconstruction network through a calibrated differentiable forward model to achieve broad generalization without retraining. Incorporating empirical scattering kernels, physics-guided regularization, and a hybrid genetic-gradient initialization strategy, DeepFilters extends the PSF from 16 micron to >400 micron in clear media and enables signal recovery beyond 120 micron deep in biological tissues, validated across fixed brain slices and sea urchin embryos.