CROSS-Calib: Selective Channel Calibration for Ultrasound Reconstruction
Organizations: University of Oxford · Peking University
Abstract
Ultrasound imaging provides views of internal tissue in real time for medical diagnosis. However, ultrasound signals may be affected by interference, and correcting errors and removing transient interference while preserving valid echoes remains challenging. We propose CROSS-Calib, a channel recovery method that uses predictions from other channels as references for amplitude and phase calibration, while a gate using context selectively combines calibrated measurements with predicted echoes to retain usable samples and repair unreliable ones. On the in vivo test set, CROSS-Calib outperforms all baseline models under gain errors and transient interference, while phantom tests show that the model restores cyst contrast with echo widths of wire targets close to the clean reference. CROSS-Calib can perform channel calibration and selective repair before beamforming to improve echo usability and image quality.
Figures & tables
| Gain | Burst | |||
| Method | A3 | A11 | A3 | A11 |
| Heavy | ||||
| No repair | 0.6873 | 0.7426 | 0.8884 | 0.9595 |
| RF-inpainting | 0.9083 | 0.9315 | 0.9958 | 0.9989 |
| IRLS-CG | 0.6627 | 0.7192 | 0.8363 | 0.8667 |
| Detect–mask CG | 0.5090 | 0.5577 | 0.6562 | 0.6812 |
| Fault | Analytical readout | Gain correction alone | CROSS-Calib |
|---|---|---|---|
| none | 0.9998/0.9998 | 1.0000 / 1.0000 | 1.0000 / 1.0000 |
| dead | 0.9353/0.7883 | 0.9394 / 0.7969 | 0.9394 / 0.7969 |
| gain | 0.9977/0.9355 | 0.9996 /0.9471 | 0.9996 / 0.9500 |
| burst | 0.9845/0.8799 | 0.9595/0.5452 | 0.9995 / 0.9970 |
| Fault | Method | gCNR | CNR (dB) | CR (dB) | Lateral FWHM (mm) | Axial FWHM (mm) |
|---|---|---|---|---|---|---|
| – | Clean reference | 0.929 | 3.90 | 19.95 | 0.515 | 0.408 |
| gain | No repair | 0.791/0.306 | 2.38/-6.08 | 12.69/3.42 | 0.514/0.522 | 0.408/0.408 |
| RF-inpainting | 0.871/0.599 | 3.22/0.15 | 15.88/8.29 | 0.516/0.524 | 0.407/0.407 | |
| IRLS-CG | 0.798/0.351 | 2.49/-4.61 | 12.87/4.04 | 0.495/0.499 | 0.427/0.422 | |
| Detect–mask CG | 0.791/0.307 | 2.38/-6.09 | 12.69/3.41 | 0.534/0.526 | 0.452/0.459 | |
| J-Sense-US | 0.830/0.662 | 2.86/0.96 | 14.07/9.37 | 0.495/0.497 | 0.428/0.427 |
Ultrasound Imaging
Ultrasound imaging uses sound waves to visualize anatomy and function, with AI improving image reconstruction, diagnosis, measurement, and scan guidance.