Modular magnetic soft robotic catheter enables visualized intervention in the pancreatobiliary tree

Authors: B. Calmé, N. J. Greenidge, A. Metcalf, A. Bacchetti, G. Loza, D. Kpeglo, P. Lloyd, V. Pensabene, 

Abstract

Diagnosis and therapy in the pancreatobiliary tree are limited by access. Guidewires reach deep into the ducts but are blind, whereas cholangioscopes provide vision but are too large to readily enter the pancreatic body and tail, the cystic duct, or the gallbladder. At the millimeter scale of these ducts, every function added to a steerable catheter competes with steering for cross-section and adds bending stiffness. We present a 1.47-mm modular magnetic soft robotic catheter in which each function is built into the magnetic body, onto its surface, or onto the footprint of the camera, and we measured what each function costs in bending under bedside fields. Imaging and fiber Bragg grating (FBG) shape sensing accounted for most of this cost, whereas a compliant gripper, magnetic anchoring balloons, and a micrometer-thick ultrasound-triggered release shell together reduced the tip angle at 8 mT only from 70° to 64°. The anchoring balloon increased the gripping force about threefold without stiffening the catheter during navigation. In live pigs, the robotic catheter performed operator-initiated task-autonomous cannulation of the native papilla without sphincterotomy and was teleoperated under onboard vision 75 mm into the pancreatic duct and, without a guidewire, through the cystic duct into the gallbladder, whose lumen it imaged. Anchoring, grasping, and payload release were each demonstrated in vivo. Adding each function at a measured cost to steering allows a soft robot small enough for these ducts to intervene as well as navigate.

Explore similar work

May 24, 2026

Stiffness Optimization for Concentrated Bending in Magnetically Actuated Catheters: Maintaining Steerability under Gradient Stiffness

Achieving both efficient pushability (propulsion transmission) and proximally concentrated bending for steerability is challenging for magnetically actuated soft catheters: higher axial/bending stiffness improves force transmission but reduces steerability, whereas lower stiffness enables large, proximally concentrated bending yet increases kinking/buckling risk under compressive push loads. To address this trade-off, we propose a stiffness-optimized multi-segment magnetically actuated catheter (SO-MAC) that integrates a decoupled steering-advancement mechanism with a gradient-stiffness architecture. The SO-MAC concentrates bending about a stable proximal pivot during advancement while the distal section passively self-straightens to transmit propulsion, aided by the optimized stiffness distribution and elastic recovery of the spring backbone against friction-induced kinking/buckling. Over 0−180∘0{-}180^{\circ} combined steering and advancement, the pivot remained stable and the distal tip advanced near-straight toward the target direction. A 1.5 mm-diameter SO-MAC achieved up to 180∘180^{\circ} steering with a 3 mm bending radius at its 10 mm tip, with an average shape error of 1.39±0.561.39 \pm 0.56 mm and a steering-pivot error of 0.35±0.100.35 \pm 0.10 mm. Visual feedback control in a bronchial phantom further confirmed robust navigation through highly curved, bifurcating paths.
Jul 28, 2026

Modular Robotic Catheters for Endovascular Aneurysm Repair

Fenestrated/Branched endovascular aneurysm repair (FEVAR/BEVAR) require surgeons to navigate catheters and guidewires into various branches of the abdominal aorta, before deploying stent grafts to alleviate pressure on the aneurysm. Previous clinical studies suggests that surgeons continue to struggle with vessel access using standard commercial instruments, prolonging the procedural time and inducing further complications. In this work, we present two contributions to solving this problem: 1) A bespoke 2-segment steerable catheter, consisting of 4 degrees of freedom to enhance dexterity. 2) An expandable, modular tendon-driven actuation platform that can accommodate for the redundancies introduced in our system. To fabricate the catheter, we capitalized on thermal fiber drawing, a technique that creates high-aspect ratio devices at scale, and processed the catheter with laser micro-machining to soften its tip. We evaluated the system using simulations, where we investigated the catheter's bending stiffness, then its steerability with in-vitro experiments in vascular phantoms. This handheld, robotic steerable catheter system has the potential to shorten the length of future endovascular surgeries, and give clinicians the tools to resolve challenging clinical cases.
Mar 4, 2026

Modeling and Control of a Pneumatic Soft Robotic Catheter Using Neural Koopman Operators

Catheter-based interventions are widely used for the diagnosis and treatment of cardiac diseases. Recently, robotic catheters have attracted attention for their ability to improve precision and stability over conventional manual approaches. However, accurate modeling and control of soft robotic catheters remain challenging due to their complex, nonlinear behavior. The Koopman operator enables lifting the original system data into a linear "lifted space", offering a data-driven framework for predictive control; however, manually chosen basis functions in the lifted space often oversimplify system behaviors and degrade control performance. To address this, we propose a neural network-enhanced Koopman operator framework that jointly learns the lifted space representation and Koopman operator in an end-to-end manner. Moreover, motivated by the need to minimize radiation exposure during X-ray fluoroscopy in cardiac ablation, we investigate open-loop control strategies using neural Koopman operators to reliably reach target poses without continuous imaging feedback. The proposed method is validated in two experimental scenarios: interactive position control and a simulated cardiac ablation task using an atrium-like cavity. Our approach achieves average errors of 2.1 +- 0.4 mm in position and 4.9 +- 0.6 degrees in orientation, outperforming not only model-based baselines but also other Koopman variants in targeting accuracy and efficiency. These results highlight the potential of the proposed framework for advancing soft robotic catheter systems and improving catheter-based interventions.