DoseRAD2026 Challenge dataset: AI accelerated photon and proton dose calculation for radiotherapy
Organizations: Department of Radiation Oncology, LMU University Hospital, LMU Munich, Munich, Germany · Division of Medical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany · National Center for Radiation Research in Oncology (NCRO), Heidelberg Institute for Radiation Oncology (HIRO), Heidelberg, Germany · GSI Helmholtzzentrum für Schwerionenforschung, Biophysics Department, Darmstadt, Germany ·
Abstract
Purpose: Accurate dose calculation is essential in radiotherapy for precise tumor irradiation while sparing healthy tissue. With the growing adoption of MRI-guided and real-time adaptive radiotherapy, fast and accurate dose calculation on CT and MRI is increasingly needed. The DoseRAD2026 dataset and challenge provide a public benchmark of paired CT and MRI data with beam-level photon and proton Monte Carlo dose distributions for developing and evaluating advanced dose calculation methods. Acquisition and validation methods: The dataset comprises paired CT and MRI from 115 patients (75 training, 40 testing) treated on an MRI-linac for thoracic or abdominal lesions, derived from the SynthRAD2025 dataset. Pre-processing included deformable image registration, air-cavity correction, and resampling. Ground-truth photon (6 MV) and proton dose distributions were computed using open-source Monte Carlo algorithms, yielding 40,500 photon beams and 81,000 proton beamlets. Data format and usage notes: Data are organized into photon and proton subsets with paired CT-MRI images, beam-level dose distributions, and JSON beam configuration files. Files are provided in compressed MetaImage (.mha) format. The dataset is released under CC BY-NC 4.0, with training data available from April 2026 and the test set withheld until March 2030. Potential applications: The dataset supports benchmarking of fast dose calculation methods, including beam-level dose estimation for photon and proton therapy, MRI-based dose calculation in MRI-guided workflows, and real-time adaptive radiotherapy.
Figures & tables
| Thorax | Abdomen | |||
| Train | Test | Train | Test | |
| Manufacturer | ViewRay | |||
| Model | MRIdian | |||
| Field strength (T) | 0.35 | |||
| Sequence | balanced steady-state free-precession (bSSFP, TrueFISP) | |||
| Acquisition | 2D, breath-hold, no contrast | |||
| Thorax | Abdomen | |||
| Train | Test | Train | Test | |
| Manufacturer | Canon Medical Aquilion/LB (97%) Siemens SOMATOM Drive (3%) | Canon Medical Aquilion/LB | Canon Medical Aquilion/LB | Canon Medical Aquilion/LB |
| kVp | 120 | 120 | 120 | 120 |
| Exposure time (ms) | 500 | 500 | 500 (94%) 750 (6%) | 500 |
| Tube current (mA) | 40 – 305 | 40 – 140 | 40 – 440 | 40 – 190 |
| Rows | 512 | 512 | 512 | 512 |
| Region | Training | Testing | External Testing | Total |
| Thoracic | 39 | 14 | 7 | 60 |
| Abdominal | 36 | 26 | – | 62 |
| Total | 75 | 40 | 7 | 122 |
| E (MeV) | (MeV) | (mm) | E (MeV) | (MeV) | (mm) | E (MeV) | (MeV) | (mm) |
| 31.73 | 6.11 | 7.49 | 114.54 | 1.47 | 5.85 | 162.93 | 0.99 | 4.49 |
| 36.80 | 5.19 | 7.62 | 116.53 | 1.44 | 5.78 | 164.45 | 0.98 | 4.46 |
| 41.38 | 4.56 | 7.70 | 118.49 | 1.42 | 5.70 | 165.96 | 0.97 | 4.44 |
| 45.60 | 4.09 | 7.77 | 120.43 | 1.39 | 5.63 | 167.46 | 0.96 | 4.41 |
| 49.54 | 3.74 | 7.82 | 122.34 | 1.37 | 5.56 | 168.95 | 0.95 | 4.39 |
| 53.25 | 3.45 | 7.86 | 124.23 | 1.34 | 5.50 | 170.43 | 0.94 | 4.36 |
| Component | Repository URL |
| Pre-processing scripts | https://github.com/DoseRAD2026/preprocessing |
| Geant4 simulation scripts | https://github.com/DoseRAD2026/geant4-dose-sim |
| Geant4 toolkit version | https://geant4.web.cern.ch/download/11.0.3.html |
| matRad VMAT planning scripts | https://github.com/DoseRAD2026/matrad-vmat-example |
| matRad commit used in this study | https://github.com/e0404/matRad/tree/71c95e96a1ecc7c7e025b9d75240ea009bdbc063 |
| pyRadPlan pencil beam baseline | https://github.com/DoseRAD2026/pyradplan-pb-baseline |
Appendix figures & tables3 assets
Supplementary material from the paper’s appendix.
Appendix
| Parameter | Description | Value |
| disp_hwd | The displacement range in voxels of the discretised search space | 6 |
| grid_sp | The grid spacing in voxels used for the effective resolution of features and deformable transformation | 5 |
| convex_mind_r | The MIND radius in voxels | 2 |
| convex_mind_d | The MIND dilation in voxels | 2 |
| adam_mind_r | The MIND radius in voxels | 1 |
| adam_mind_d | The MIND dilation in voxels | 2 |
| Field | Unit | Description |
| Per beam | ||
| beam_idx | – | Sequential beam index (0-indexed) |
| iso_center | mm | 3D coordinate of the beam rotation center in patient |
| num_mlc_leaf_pairs | – | Number of leaf pairs to shape the segment |
| control_points | – | Discrete beam delivery sampling points along the arc |
| Per control point | ||
| Field | Unit | Description |
| iso_center | mm | 3D coordinate of the beam rotation center in patient |
| Per beam | ||
| beam_idx | – | Sequential beam index (0-indexed) |
| gantry_angle | deg | Angle of the proton beam direction around patient |
| rays | – | A group of pencil beamlets from the same BEV grid position within the beam |
| Per ray | ||