Acquisition state behaves as a structured, measurable variable governing lung-nodule AI: kernel-driven measurement instability and noise-driven detection fragility, invisible to DICOM metadata

Authors: Daniel Soliman

Abstract

AI governance for medical imaging is formalizing: the 2026 ACR-SIIM Practice Parameter recommends local acceptance testing and ongoing drift monitoring, and the ACR Assess-AI registry monitors AI outputs using DICOM metadata for context. We argue that a necessary, currently unmonitored layer sits beneath output metrics: whether incoming studies remain within the acquisition envelope a model was validated on. Using a LUNA16-trained MONAI RetinaNet lung-nodule detector, we test whether acquisition state behaves as a structured, measurable variable. On real paired CT differing only in reconstruction kernel (NLST B30f vs B80f), kernel alone shifted AI-measured diameter and flipped a Fleischner size category in 5.2% (8 of 155) of nodules at fixed patient and acquisition, while detection confidence was unchanged (Wilcoxon p=0.22). Under controlled LIDC-IDRI perturbations the effects dissociated by axis: the noise axis degraded detection confidence (p=5.9e-32, concentrated in nodules under 6 mm) but not measurement, while the frequency/kernel axis corrupted measurement (p=8.6e-13) but not detection. A 4-feature pixel fingerprint recovered reconstruction identity (patient-level AUC about 0.95 on real CT, 0.995 on a QIBA phantom) where the ConvolutionKernel DICOM tag was uninformative (identical labels across reconstructions). The kernel axis transported across four manufacturers (leave-one-vendor-out AUC 0.94-0.98, matching the within-vendor ceiling). Acquisition state thus maps to distinct AI failure modes, frequency content to measurement reliability and noise to detection sensitivity, and is not recoverable from metadata. Acquisition-aware, input-side validation is the missing layer for the acceptance-testing and drift-monitoring requirements now entering imaging-AI accreditation.

Explore similar work

May 1, 2026eess.IV

Reconstruction Interval Z-Phase Dependence of AI Detection Sensitivity in CT Lung Nodule Screening

Background: Sensitivity of AI-assisted lung nodule detection systems is known to vary with CT acquisition parameters including radiation dose, reconstruction kernel, and slice thickness. However, the dependence of detection probability on nodule position within the reconstruction cycle -- the z-phase -- has not, to the author's knowledge, been characterized for deep learning-based detection systems. Methods: A retrospective analysis was performed using the LIDC-IDRI dataset. Detection results from a previously validated 154-case perturbation study were re-analyzed. For each consensus nodule (>=4-reader agreement), z-phase was defined as the fractional position of the nodule center within the reconstruction cycle, folded to [0, 0.5]. Detection sensitivity was stratified by z-phase bin, reconstruction interval (1mm, 3mm, 5mm), and by the ratio of reconstruction interval to nodule diameter (d/D). Results: At 5mm reconstruction interval, sensitivity was 71.6% vs 84.8% at 1mm baseline. Within the 5mm condition, sensitivity varied by 17.6 percentage points across z-phase bins. Stratified by d/D ratio, sensitivity was 92.4% for d/D < 0.5, 78.0% for 0.5 <= d/D < 1.0, and 61.4% for d/D >= 1.0, with a systematic z-phase effect present only in the d/D >= 1.0 stratum. Conclusions: AI detection sensitivity depends on the ratio of reconstruction interval to nodule diameter. When this ratio approaches or exceeds 1.0 -- as occurs for 3-6mm nodules at 5mm reconstruction -- z-phase becomes the dominant source of per-study detection variance. This stochastic effect is invisible to protocol-level quality metrics and not reflected in AI confidence scores.
Dan Soliman
May 31, 2026eess.IV

Sparse-View Lung Nodule Volumetry from Digitally Reconstructed Radiographs via AReT: Anatomy-Regularized TensoRF

We identify and resolve a previously unreported failure mode in TensoRF when applied to X-ray attenuation fields: the default density shift of -10, originally introduced for RGB scene reconstruction, suppresses density gradients and prevents sparse-view medical reconstruction regardless of learning rate or regularization strategy. Setting the density shift to zero restores gradient flow and enables stable volumetric reconstruction of pulmonary nodules from only three orthogonal X-ray projections. Building on this, we propose AReT, an anatomy-regularized tensorial radiance field framework for lung nodule reconstruction using coronal, sagittal, and axial projections from the LIDC-IDRI dataset (19 patients, radiologist-annotated nodules). Unlike existing NeRF approaches requiring dense multi-view acquisition, AReT is designed for sparse-view thoracic imaging and incorporates chest-anatomy-aware regularization combining L1 sparsity and total variation smoothness. A systematic comparison across 11 reconstruction strategies shows anatomy-aware regularization consistently outperforms generative-prior-guided approaches. Evaluated against radiologist consensus segmentations, AReT achieves Pearson r=0.983 (p<0.0001) for clinically actionable nodules >=10 mm (n=14), median absolute volumetric error of 11.4%, near-zero systematic bias of -77.3 mm^3, and 8.4x improvement over spherical volume approximation.
Spoorthi M, Suja Palaniswamy
May 14, 2026cs.AI

How Sensitive Are Radiomic AI Models to Acquisition Parameters?

A main barrier for the deployment of AI radiomic systems in clinical routine is their drop in performance under heterogeneous multicentre acquisition protocols. This work presents a performance-oriented framework for quantifying scan parameter sensitivity of radiomic AI models, while identifying clinically significant parameter regions associated with improved cross-dataset robustness. We formulate a mixed-effects framework for quantifying the influence that clinically relevant acquisition parameters have on models performance, while accounting for subject-level random effects. We have applied our framework to lung cancer diagnosis in CT scans using two independent multicentre datasets (a public database and own-collected data) and several SoA architectures. To evaluate across-database reproducibility, CT parameters have been adjusted using the data collected and tested on the public set. The optimal configuration selected is the current of the X-ray tube >= 200 mA, spiral pitch <= 1.5, slice thickness <= 1.25 mm, which balances diagnostic quality with low radiation dose. These configuration push metrics from 0.79+-0.04 sensitivity, 0.47+-0.10 specificity in low quality scans to 0.90+-0.10 sensitivity, 0.79 +- 0.13 specificity in high quality ones.
D. Gil, I. Sanchez, C. Sanchez