CardioFusion-AI: Robust ECG--PPG Fusion for Multimodal Physiological Monitoring Under Signal Degradation
Organizations: Department of Electronics and Communication Engineering, Saveetha School of Engineering · Government Kilpauk Medical College and Hospital
Abstract
Wearable electrocardiogram (ECG) and photoplethysmogram (PPG) sensors are complementary but individually fragile: motion artifact, poor contact, and sensor dropout can degrade one or both signals. Fusion strategies that assume both modalities are equally trustworthy can become less reliable than a single clean modality under degradation. We present CardioFusion-AI, a framework whose signal-processing front end, including R-peak and systolic-peak detection, an Orphanidou-type signal-quality index, and beat-by-beat pulse transit time estimation, is validated on 53 real intensive-care recordings (848 windows; heart-rate mean absolute error 1.61 bpm for ECG and 2.78 bpm for PPG) and a real annotated fetal ECG database (R-peak F1 0.89-0.98). We then conduct a controlled synthetic degradation study comparing eight ECG-PPG fusion strategies across six degradation regimes spanning graded corruption and complete modality loss, using five independent training seeds. Attention fusion achieved the lowest descriptive overall error (1.66+/-0.43 bpm). Both adaptive gates reallocated weight toward the healthy modality under complete modality loss, but showed near-zero correlation between gate weight and signal quality under graded degradation (r = 0.10-0.24). Signal-quality conditioning produced a specific improvement under missing-PPG conditions (1.56+/-0.59 bpm), approaching the 1.48 bpm unimodal ceiling. With only five training seeds, no pairwise comparison survives Holm-corrected significance testing; effect sizes and confidence intervals are therefore reported. These results indicate that modality availability and modality quality are functionally distinct problems for adaptive fusion.
Figures & tables
| Regime | ECG level | PPG level | # grid cells |
|---|---|---|---|
| Both clean | 0 | 0 | 1 |
| ECG degraded only | 1–3 | 0 | 3 |
| PPG degraded only | 0 | 1–3 | 3 |
| Both degraded | 1–3 | 1–3 | 9 |
| ECG missing | 4 | 0 | 1 |
| PPG missing | 0 | 4 | 1 |
| Model | Overall | Both clean | ECG degraded | PPG degraded | Both degraded | ECG missing | PPG missing |
|---|---|---|---|---|---|---|---|
| ECG-only | 2.18 0.45 | 1.86 0.76 | 1.52 0.41 | 1.41 0.56 | 1.43 0.41 | 14.34 0.38 | 1.48 0.67 |
| PPG-only | 2.61 0.30 | 1.49 0.32 | 1.38 0.30 | 1.92 0.22 | 2.13 0.35 | 1.47 0.34 | 15.02 1.51 |
| Fixed-average fusion | 2.63 0.64 | 2.39 0.71 | 2.11 0.89 | 2.14 0.73 | 2.17 0.76 | 7.01 0.82 | 5.77 1.33 |
| Feature-level fusion | 2.87 0.53 | 2.53 0.57 | 2.17 0.31 | 2.25 0.53 | 2.40 0.67 | 6.95 0.76 | 7.28 1.22 |
| Attention fusion | 1.66 0.43 | 1.24 0.26 | 1.65 0.39 | 1.18 0.18 | 1.55 0.31 | 3.17 1.91 | 2.97 1.73 |
| Global-weighted late fusion | 2.67 0.30 | 1.98 0.44 | 1.83 0.62 | 2.05 0.35 | 2.26 0.40 | 6.86 0.54 | 7.24 0.80 |
| Comparison | Regime | Mean diff. | |||
|---|---|---|---|---|---|
| SQI vs. Attn. | both clean | +0.74 | 2.51 | 0.063 | 1.00 |
| SQI vs. Attn. | PPG degraded | +0.63 | 1.50 | 0.063 | 1.00 |
| SQI vs. Implicit | overall | 0.07 | 0.18 | 0.81 | 1.00 |
| Implicit vs. Attn. | overall | +0.40 | 0.56 | 0.31 | 1.00 |
Multimodal Sensor Fusion
Multimodal sensor fusion combines data from different sensors to improve perception, tracking, localization, and prediction.
See all 417 papers