Surgical Master Console Using General-Purpose Robot Arms and a Separable Articulated Distal Interface: Porcine In-Vivo Evaluation

Authors: Minsung Kim, Seonho Shim, Dongho Yee, Younghoon Noh, Juahn Oh, Yechan Seo, Jinseok Lee, Jiyul Lee, +4 more

Organizations: Department of Transdisciplinary Medicine, Seoul National University Hospital, Seoul, Republic of Korea. · Rosota Inc., Seoul, Republic of Korea. · Department of Mechanical Engineering, Seoul National University, Seoul, Republic of Korea. · Department of Mechanical Engineering, Chungang University, Seoul, Republic of Korea. · Department of Computer Science and Engineering, Seoul National University, Seoul, Republic of Korea. · Eulji University College of Medicine, Daejeon, Republic of Korea. · Department of Medicine, Seoul National University College of Medicine, Seoul, Republic of Korea. · Interdisciplinary Program in Medical Informatics, Seoul National University College of Medicine. · Institute of Convergence Medicine with Innovative Technology, Seoul National University Hospital, Seoul, Republic of Korea.

Abstract

High-performance surgical master consoles offer intuitive articulated manipulation but are expensive and difficult to reproduce, while accessible commercial haptic devices lack built-in interfaces for surgical wrist articulation and continuous grasp. We present a laparoscopic master in which general-purpose robot manipulators provide the programmable base and surgical-specific interaction is concentrated in a separable distal adapter. Each 6-DoF arm carries a custom 2-DoF direct-drive wrist, continuous grasp sensing, and clutch-based workspace management; the complete bimanual hardware costs approximately US$5,500, below common bimanual commercial haptic references. To support sustained hand-guided use, we characterize current scaling, gravity, and break-away friction and use them for posture-dependent gravity compensation and motion-gated assistance, reducing continuous actuation and thermal loading. A common master state drives either Isaac Sim or an RCM-constrained dual-FR3 platform. Master-state publication to FR3 target update adds at most 8 ms, and tip tracking shows 0.31-0.66 mm mean delay-aligned residuals. In a porcine cholecystectomy experiment, five common operator-side metrics and hand/tip-speed distributions were compared descriptively with human in-vivo telemetry from 120 clinical Versius recordings of 99 cholecystectomy procedures; no interruption originated from the master console, and no abnormal clinical signs were recorded during the 7-day postoperative observation period. These results demonstrate a surgical master built from general-purpose robot arms with surgical-specific interaction concentrated in a separable distal interface.

Figures & tables

Explore similar work

Oct 5, 2026

DASH: A da Vinci Adapter for Serial-link and Humanoid Robots as an Accessible Platform for Surgical Robotics Research

Robotic minimally invasive surgery offers well-documented clinical benefits, but the cost and infrastructure requirements of purpose-built platforms limit access in rural and lower-resourced facilities. Recent work has teleoperated general-purpose robots for laparoscopic tasks and in vivo procedures, but relied on handheld instruments coupled through passive linkages rather than native robotic actuation. Instead, we adapt da Vinci Classic and Xi instruments onto general-purpose robots that can integrate in clinical workflows. We present DASH, a da Vinci Adapter for Serial-link and Humanoid platforms, consisting of two types of adapters that require no modification to the instruments themselves. Each adapter is compatible with many robotic platform load capacities, engages the instrument's native latch, and wirelessly identifies inserted tools to load instrument-specific kinematics and coupling matrices. Teleoperated experiments demonstrate the efficacy of DASH and its benefits for enabling research access to surgical robotic platforms.
Sep 8, 2026

A Controlled Comparison of Manual and Teleoperated Intraocular Instrument Motion for an Input Device

Input devices for robotic microsurgery are frequently described as preserving the surgeon's trained technique, but the claim is rarely measured. We compared manual and teleoperated intraocular instrument motion with the trocar constraint, the instrument, the eye model and the tracking source common to both conditions, so that the control interface was the only factor varied. Prior comparisons cannot hold the instrument fixed, because a robotic instrument is not the tool used manually. Sixteen participants performed a navigation task on a commercial ophthalmic simulator by hand and through a three-degree-of-freedom input device commanding a five-joint robot. Task outcome was equal but at ceiling: every participant acquired all five targets under both interfaces with no retinal or lens injury. Execution differed on every measure. Teleoperated trials took three times as long at a quarter of the median speed, covered less than half the angular working range, and were broken into 3.5 times as many separate movements. Completion time and movement fragmentation improved substantially across four trials of practice and had not plateaued; the measures set by the configured rate ceiling and joint limit changed the least. Finger activity doubled and pinch variability tripled, so reducing instrument degrees of freedom redistributed manual effort rather than reducing it. The interface preserves the outcome and reshapes the execution.
Jul 8, 2026

In vivo feasibility study of humanoid robots in surgery

Recent advances in actuation, control and learning have rapidly pushed humanoid robots from a distant vision towards near-term real-world deployment. Healthcare is a particularly pressing domain, in which staffing shortages and increasing care demand are widening the gap between clinical workload and available skilled labour. Although current automation has largely focused on digital and logistical tasks, much hospital work remains embodied, requiring mobility, manipulation and safe interaction in human-designed environments. Humanoid form factors offer unique potential, particularly for assisting with surgical tasks. Traditionally, robotic systems for surgery are purpose-built platforms such as Intuitive Surgical's da Vinci Surgical System, and it remains unclear how close current humanoid systems are to meeting the precision, control and safety requirements of minimally invasive surgery. Here we present a systematic evaluation of contemporary humanoid technology for laparoscopic surgical tasks. We develop a humanoid-based laparoscopic teleoperation framework using general-purpose instruments and assess its abilities through benchtop characterization, dry-laboratory user studies spanning diverse surgical experience levels and in vivo porcine studies. Across these evaluations, we quantify technical feasibility, task performance and clinical readiness relative to established surgical platforms. Together, our study provides an evidence-based assessment of current humanoid abilities and limitations for surgical applications, highlighting both their promise and key technical challenges that must be addressed before clinical deployment.