Continuous wavelet-based feature extraction for laser-induced breakdown spectroscopy classification of human plasma samples
Organizations: Miner School of Computer and Information Sciences, Kennedy College of Sciences, University of Massachusetts, Lowell, USA · Department of Computer Systems Technology, North Carolina Agriculture and Technical University
Abstract
Extracting discriminative features from laser-induced breakdown spectroscopy (LIBS) spectra is important for the classification of complex biological samples. This study evaluates continuous wavelet transform (CWT)-based spectral peak extraction for distinguishing human plasma samples labeled as SARS-CoV-2-positive or negative. Plasma specimens from 181 individuals, comprising 53 positive samples and 128 negative controls, were deposited as 5 μL droplets on silicon substrates and dried before analysis. Spectra were acquired using a nanosecond Nd:YAG laser operating at 1064 nm, with 100 single-shot measurements collected per specimen and averaged after intensity-based outlier rejection. CWT-based peak detection was coupled with a support vector machine (SVM) classifier. The reported leave-one-out evaluation yielded an accuracy of 93.00%, sensitivity of 94.67%, and specificity of 91.61%; these estimates remain provisional pending reconciliation with specimen-level predictions. Features identified by the CWT pipeline included emission lines assigned to Ca II, K I, and N I. The workflow provides a basis for examining spectral features associated with the studied plasma groups. Independent validation and assessment of specimen, substrate, and acquisition-related variability are required to establish generalizability and diagnostic relevance.
Figures & tables
| Plasma specimen collection Round 1: 118 specimens (24 positive, 94 negative) Round 2: 63 specimens (29 positive, 34 negative) |
| Combined LIBS cohort 181 individuals: 53 positive, 128 negative controls |
| Specimen preparation 5 L plasma on silicon; infrared drying for 10 min |
| LIBS acquisition 1064 nm Nd:YAG laser; 100 single-shot spectra per specimen |
| Method | Accuracy (%) | Sensitivity (%) | Specificity (%) |
|---|---|---|---|
| CWT-SVM | 93.00 | 94.67 | 91.61 |
| Excluded wavelength (nm) | Accuracy (%) | Sensitivity (%) | Specificity (%) | Accuracy decrease (pp) |
|---|---|---|---|---|
| None | 93.00 | 94.67 | 91.61 | 0.00 |
| 396.880 | 92.42 | 91.65 | 89.46 | 0.58 |
| 288.179 | 87.45 | 82.34 | 84.87 | 5.55 |
| 413.085 | 88.23 | 84.67 | 83.65 | 4.77 |
| 393.398 | 84.65 | 82.13 | 80.91 | 8.35 |
Medical Diagnosis
Medical diagnosis uses clinical data, tests, and images to identify diseases, assess risk, and guide care.
See all 277 papers