Feasibility of Multifrequency Neural Network-Based Wave Inversion in Ultrasound Time-Harmonic Elastography
Authors: Héloïse Bustin, Tom Meyer, Paul Spiesecke, Jakob Jordan, Anja Hennemuth, Ingolf Sack
Organizations: Deutsches Herzzentrum der Charité, Institute of Computer-assisted Cardiovascular Medicine, Augustenburger Platz 1, 13353 Berlin, Germany · Charité – Universitätsmedizin Berlin, corporate member of Freie Universität Berlin and Humboldt Universität zu Berlin, Charitéplatz 1, 10117 Berlin, Germany · Fraunhofer MEVIS, Berlin, Germany · DZHK (German Centre for Cardiovascular Research), Berlin, Germany
Ultrasound time-harmonic elastography (THE) is a non-invasive imaging technique that measures soft-tissue stiffness. Relying on the same principle of time-harmonic shear wave excitation as established magnetic resonance elastography (MRE), THE offers full field-of-view stiffness mapping at short acquisition times and sub-second temporal resolution across organs. However, reconstruction of shear wave speed maps from acquired tissue wavefields remains challenging due to speckle noise and sparse spatial coverage of the lateral component. Deep learning methods have shown promising results for image reconstruction in transient-based ultrasound elastography modalities but remain unexplored in THE. In MRE, MF-ElastoNet, a 2D multifrequency neural network-based wave reconstruction method trained on synthetic physics-based wave data, has shown high generalizability across dynamic ranges and resolutions. We investigated the potential of a multifrequency neural network-based wave inversion for THE, by performing a first feasibility analysis. To this end, we adapted the processing pipeline of MF-ElastoNet from MRE to THE. We tested our approach on THE acquired phantom data from the algorithms for mapping elastic modulus Challenge, as well as on in vivo liver THE in 4 healthy volunteers and 14 patients with focal liver lesions, including benign focal nodular hyperplasia, and malignant hepatocellular carcinoma and adenocarcinoma. MF-ElastoNet was successfully adapted to THE with in vivo liver testing. Compared with k-MDEV, the state-of-the-art wave inversion method for THE, Bland-Altman analysis showed a mean bias of -0.02 m/s (95% limits of agreement: -0.36 to 0.33 m/s).
Figures & tables
Figure 1: Wave inversion processing pipeline: A) Multifrequency complex-valued wavefields were smoothed. 12 × 12 patches were then extracted from the images and downsampled to 4 × 4. B) Multifrequency 4 × 4 complex-valued patches were processed by MF-ElastoNet, obtaining multifrequency SWS (shear wave speed) and quantifying the associated prediction uncertainty via the CV (coefficient of variation). C) Processed SWS values and associated uncertainty were mapped to frequency-resolved SWS and CV maps. These were averaged to obtain the final SWS and CV maps.
Figure 2: Wavefields, SWS (shear wave speed), and CV (coefficient of variation) maps predicted by MF-ElastoNet for phantoms I, II, and III as well as their respective ROIs (regions of interest) as a white dashed line.
Phantom
Metric
k-MDEV
MF-ElastoNet
Phantom I
SWS std. (m/s)
0.083
0.178
Phantom II
CNR
2.87
1.57
Edge width (mm)
8.16
3.66
Edge accuracy (mm)
-2.37
1.25
Phantom III
CNR
2.09
2.69
Edge width (mm)
2.93
6.55
Table 1: aMEM Challenge metrics for k-MDEV and MF-ElastoNet: For phantom I, the spatial variation of the SWS (shear wave speed) is reported through the standard deviation (std.). For phantoms II and III, the CNR (contrast-to-noise ratio), edge sharpness through the edge width, and the edge accuracy are reported.
Figure 3: B-mode, wavefield, SWS (shear wave speed) maps produced by k-MDEV and MF-ElastoNet, and CV (coefficient of variation) maps predicted by MF-ElastoNet are reported for two healthy volunteers, and three patients, including benign (23-year-old female with focal nodular hyperplasia) and malignant (81-year-old female with hepatocellular carcinoma and 49-year-old male with metastatic colorectal adenocarcinoma) cases. Liver parenchyma (white) and tumor (magenta) are additionally shown overlaid on all images.
Figure 4: Bland-Altman analysis of SWS (shear wave speed) values: MF-ElastoNet vs k-MDEV. Color denotes the subject groups (green: healthy volunteer, blue: focal nodular hyperplasia (FNH), purple: hepatocellular carcinoma (HCC), brown: adenocarcinoma), and marker shape tissue type (circle: liver parenchyma; cross: tumor).
Ultrasound imaging uses sound waves to visualize anatomy and function, with AI improving image reconstruction, diagnosis, measurement, and scan guidance.
Departments of Radiology, Biomedical Engineering, Electronic and Computer Engineering, and Mechanical Engineering, University of Iowa, Iowa City, IA 52246 · Department of Radiology, University of Iowa, Iowa City, IA, 52246
Department of Mathematics, The University of Utah, address USA. · Department of Mathematics, Virginia Tech, address USA. · Department of Mathematics, The University of Utah, address, address, USA.