Real-time Monocular 2D and 3D Perception of Endoluminal Scenes for Controlling Flexible Robotic Endoscopic Instruments
Authors: Ruofeng Wei, Kai Chen, Yui Lun Ng, Yiyao Ma, Justin Di-Lang Ho, Hon Sing Tong, Xiaomei Wang, Jing Dai, +2 more
Organizations: Department of Computer Science and Engineering, The Chinese University of Hong Kong, Hong Kong, China. · Department of Mechanical Engineering, The University of Hong Kong, China. · Department of Mechanical and Automation Engineering, The Chinese University of Hong Kong, and also with the Department of Mechanical Engineering, The University of Hong Kong.
Endoluminal surgery offers a minimally invasive option for early-stage gastrointestinal and urinary tract cancers but is limited by surgical tools and a steep learning curve. Robotic systems, particularly continuum robots, provide flexible instruments that enable precise tissue resection, potentially improving outcomes. This paper presents a visual perception platform for a continuum robotic system in endoluminal surgery. Our goal is to utilize monocular endoscopic image-based perception algorithms to identify position and orientation of flexible instruments and measure their distances from tissues. We introduce 2D and 3D learning-based perception algorithms and develop a physically-realistic simulator that models flexible instruments dynamics. This simulator generates realistic endoluminal scenes, enabling control of flexible robots and substantial data collection. Using a continuum robot prototype, we conducted module and system-level evaluations. Results show that our algorithms improve control of flexible instruments, reducing manipulation time by over 70% for trajectory-following tasks and enhancing understanding of surgical scenarios, leading to robust endoluminal surgeries.
Figures & tables
Fig. 1: Robotic instrument configurations and endoscopic scenarios in continuum robotic endoluminal surgery. (a) For transurethral bladder tumour resection, the robotic instruments are deployed through a standard urology outer sheath alongside a standard telescope. (b) Two external instrument channels equipped to a standard single-channel GI endoscope (e.g. from Olympus) to deliver the robotic instruments. (c) Monocular endoluminal image captured by GI endoscope.
Fig. 2: Overview of the proposed image-based perception framework for continuum robotic endoluminal surgery. The framework consists of three modules in its hybrid training: 1) 2D continuum robot segmentation module for flexible instruments recognition; 2) 3D robot state estimation module for flexible instrument shape calculation ; and 3) 3D depth estimation module for geometric information provision about the entire endoscopic scene. Monocular endoscopic images serve as the sole input for the perception framework. After training, the framework is deployed on a novel continuum robotic system.
Fig. 3: Calculation of the Per-Pixel Rendering ( PPR ) field from the depth estimation, which is the process of PPR calculation illustrated in Fig. 2 . The PPR field shows a strong correlation with the corresponding input endoscopic image. Arrows from Depth indicate how depth is transformed into attenuation and light direction. Additionally, arrows between Pseudo and Real images represent the correlation computation, with the correlation values normalized to a range of 0 to 1.
Fig. 4: Monocular depth estimation model for continuum robotic endoluminal surgical scene.
Fig. 5: Physically-realistic simulation of continuum robotic system for endoluminal surgery. (a) Setup of the continuum robotic system. (b) Configuration of flexible instruments approximated by curve segments in camera coordinate system. The flexible robot consists of three segments: Shoulder OS , Proximal SP , and Distal PD . α and β denote internal bending angles of SP and PD . γ and δ represent OS yaw and roll angles. (c) Comparison of real and synthetic endoscopic images. More comparison results are shown in Video 1.
Fig. 6: Overview of the realistic synthetic continuum robot surgical data generation framework. More synthetic datasets can be found in Video 2.
Fig. 7: 3D texture mapping pipeline for extracting high-quality and continuous textures from endoscopic images and depth information, which is the process in Fig. 6 .
Fig. 8: Examples of synthetic continuum robot surgical data with multiple robot configurations and endoscope scenes.
Method
Left
Right
Dice ↑
IoU ↑
Precision ↑
Dice ↑
IoU ↑
Precision ↑
Syn
38.70
27.93
68.86
15.87
12.09
36.62
Copy-Paste
60.82
48.02
79.34
29.49
20.04
74.74
Ours
94.93
90.38
94.65
95.59
91.60
95.04
TABLE I: Quantitative comparisons for 2D flexible instrument segmentation on ex-vivo data. ↑ indicates that higher values of the corresponding metric represent better performance.
Fig. 9: Performance of flexible instrument segmentation tracking. (a) Examples of instrument segmentation visualization using different approaches during tracking. The semitransparent shadows indicate the predicted segmentations of the instruments. Our results align more accurately with the robotic configuration. (b) Precision comparison of the Ours , Copy-Paste [ 70 ] , and Syn segmentation methods against the ground truth. The numbered circles (1, 2, 3) indicate specific time points that align with the correspondingly numbered images in (a). Values represent the average precision across both left and right instrument segmentations.
Metrics
KP
SKL
Quat
Rot6D
Ours
α
Mean (∘)↓
22.46
20.61
7.08
4.16
2.60
Med (∘)↓
24.45
22.70
3.09
2.43
2.08
β
Mean (∘)↓
25.82
19.50
7.11
5.54
5.34
Med (∘)↓
24.51
19.34
5.24
3.83
3.95
γ
Mean (∘)↓
29.89
17.70
8.31
7.09
6.66
Med (∘)↓
26.34
13.68
6.74
5.44
5.51
TABLE II: Quantitative comparison of our methods with the state-of-the-art methods on testing datasets.
Fig. 10: Qualitative comparison of our continuum robot state estimation model with the state-of-the-art methods: keypoint-based KP [ 28 ] , skeleton-based SKL [ 73 ] , quaternion pose representation Quat [ 33 ] , and rotation 6D pose representation Rot6D [ 34 ] . First, the flexible instruments were manipulated using the estimated state parameters within the proposed simulator. Then, the simulated instruments were projected onto 2D coordinates system to render synthetic images, utilizing the camera extrinsic and intrinsic matrices.
Fig. 11: Performance of continuum robot state tracking. (a) Comparison examples between real and rendered endoscopic images. (b) Variations in tracking error across all estimated robot states over time. The time corresponding to the second frame in (a) with high error is highlighted with a red circle, and its error and uncertainty are presented.
Fig. 12: Qualitative comparison of monocular depth estimation on synthetic and real endoscopic data. The depth estimation results of our model are compared with foundation models Metric3D [ 76 ] and EndoOmni [ 43 ] , finetuned Depth Anything (DA- finetuned ) [ 37 ] , ZoeDepth [ 77 ] and ResNet [ 69 ] . Each row shows flexible robots in the same configuration. Our method captures depth with fine details, where red indicates greater distance from the endoscope and blue indicates proximity.
Metrics
Definition
Abs Rel
∣D∣1∑d∈D∣d∗−d∣/d∗
Sq Rel
∣D∣1∑d∈D∣d∗−d∣2/d∗
RMSE
∣D∣1∑d∈D∣d∗−d∣2
RMSElog
∣D∣1∑d∈D∣logd∗−logd∣2
η
∣D∣1{d∈D∣max(dd∗,d∗d<υ)∣}×100%
TABLE III: Depth Evaluation Metrics
Method
Finetuning
Supervision
Error↓
Accuracy↑
Abs Rel
Sq Rel
RMSE
RMSElog
η<1.251
η<1.252
η<1.253
Metric3D [ 76 ]
✗
−
0.239
3.671
12.543
0.298
0.494
0.847
0.986
EndoOmni [ 43 ]
✗
−
0.273
4.048
11.370
0.354
0.474
0.769
0.92
ResNet [ 69 ]
✓
GT
0.169
1.490
7.703
0.255
0.712
0.917
0.965
ZoeDepth [ 77 ]
✓
GT
0.114
0.857
6.210
0.174
0.867
0.953
0.999
Depth Anything [ 37 ]
✓
GT
0.113
0.806
5.094
0.160
0.858
0.952
0.995
TABLE IV: Quantitative comparisons of monocular depth estimation on synthetic robotic endoluminal test data. GT, ground truth; SSL, self-supervised learning. The best results are in bold. All values in mm .
Fig. 13: Reconstructed 3D point cloud from the monocular depth estimation of real continuum robotic endoluminal surgical data. Variable i denotes the frame index in the video sequence. More results of monocular depth estimation can be found in Video 3.
Lsup
PRMod
Error↓
Accuracy↑
Abs Rel
Sq Rel
RMSE
RMSElog
η<1.251
✗
✗
0.096
0.718
4.864
0.150
0.842
✓
✗
0.088
0.610
4.576
0.145
0.870
✗
✓
0.093
0.664
4.728
0.146
0.858
✓
✓
0.072
0.395
4.286
0.102
0.945
TABLE V: Ablation study on Lsup and PRMod .
Fimg
FPPR
Error↓
Accuracy↑
Abs Rel
Sq Rel
RMSE
RMSElog
η<1.251
✓
✗
0.074
0.439
4.486
0.107
0.936
✗
✓
0.073
0.399
4.323
0.105
0.943
✓
✓
0.072
0.395
4.286
0.102
0.944
TABLE VI: Ablation study on image features Fimg and PPR features FPPR .
Modules
Execution time (ms)
2D segmentation
5.50
3D state estimation
20.72
Subtotal (sequential)
26.22
3D depth estimation (parallel)
47.15
Total system latency
47.15
TABLE VII: Runtime performance of monocular image-based 2D and 3D perception framework.
Fig. 14: Flexible instruments 3D shape monitoring. (a) Examples of endoscopic views alongside the corresponding robotic shapes visualized in the software. The 3D shapes of two instruments are shown from both top and front views. (b) Uncertainty in robot state estimation over time, with fives shapes from (a) marked. Areas with uncertainty values greater than 1 are overlaid with a semi-transparent orange color.
Fig. 15: Failure cases in robot state estimation. Instruments marked with dashed ovals illustrate how partial visibility results in incorrect state estimation and increased uncertainty.
Fig. 16: Analysis of instrument-tissue deviation during flexible instrument touching. (a) Four representative frames along with their corresponding depth maps during the contact events. (b) Variation in distance as the flexible instrument approaches the tissue twice. (c) Multiple views of the 3D structures during the touching events. One demonstration can be found in Video 4.
Department of Mechanical and Automation Engineering and T Stone Robotics Institute, The Chinese University of Hong Kong, Hong Kong · School of Mechanical and Electronic Engineering, Wuhan University of Technology, Wuhan 430070, China · state Key Laboratory of Intelligent Manufacturing Equipment and Technology, Huazhong University of Science and Technology (HUST), Wuhan, China +1