sFRC for assessing hallucinations in medical image restoration

Authors: Prabhat Kc, Rongping Zeng, Nirmal Soni, Aldo Badano

Organizations: Center for Devices and Radiological Health in FDA, Silver Spring, MD, 20993, USA

Abstract

Deep learning (DL) methods are currently being explored to restore images from sparse-view-, limited-data-, and undersampled-based acquisitions in medical applications. Although outputs from DL may appear visually appealing based on likability/subjective criteria (such as less noise, smooth features), they may also suffer from hallucinations. This issue is further exacerbated by a lack of easy-to-use techniques and robust metrics for the identification of hallucinations in DL outputs. In this work, we propose performing Fourier Ring Correlation (FRC) analysis over small patches and concomitantly (s)canning across DL outputs and their reference counterparts to detect hallucinations (termed as sFRC). We describe the rationale behind sFRC and provide its mathematical formulation. The parameters essential to sFRC may be set using predefined hallucinated features annotated by subject matter experts or using imaging theory-based hallucination maps. We use sFRC to detect hallucinations for three undersampled medical imaging problems: CT super-resolution, CT sparse view, and MRI subsampled restoration. In the testing phase, we demonstrate sFRC's effectiveness in detecting hallucinated features for the CT problem and sFRC's agreement with imaging theory-based outputs on hallucinated feature maps for the MR problem. Finally, we quantify the hallucination rates of DL methods on in-distribution versus out-of-distribution data and under increasing subsampling rates to characterize the robustness of DL methods. Beyond DL-based methods, sFRC's effectiveness in detecting hallucinations for a conventional regularization-based restoration method and a state-of-the-art unrolled method is also shown.

Figures & tables

Appendix figures & tables20 assets

Supplementary material from the paper’s appendix.

Appendix

Explore similar work

Aug 6, 2026

Does FLAIR super-resolution erase or hallucinate small white-matter lesions?

White matter hyperintensities (WMH), bright regions on Fluid-attenuated Inversion Recovery (FLAIR) scans are associated with cerebrovascular pathology and neurodegeneration. FLAIR is usually acquired with thick slices in clinical settings, giving it poor through-plane resolution. Super-resolution (SR) is a widely used method for recovering an isotropic volume from an anisotropic scan. Yet whether applying it prior to WMH segmentation preserves lesion content remains unknown: a model may erase small real lesions or hallucinate absent ones. We used 1-mm isotropic high-resolution (HR) FLAIR scans from 29 individuals in the ADNI cohort, each manually segmented for WMH by an expert. Then, we degraded each to simulated 3 and 5 mm through-plane acquisitions. Multi-contrast implicit neural representation (INR), a single-contrast self-supervised model (ECLARE), and cubic interpolation were used to upsample them onto the HR grid. WMH segmentation from a simulated thick slice and the original HR FLAIR set the floor and ceiling, respectively, for the per-lesion analysis. Of four WMH segmentation methods (WMH-SynthSeg, segcsvd, MARS-WMH, TrUE-Net), we ran the analysis under the most sensitive one to small lesions on HR (MARS-WMH) with the evaluation metrics of detection sensitivity, erasure rate (HR-detected lesions lost after reconstruction), and hallucination rate (predicted components absent from both the manual and HR segmentation). The dominant effect of SR was erasure of small real lesions, not hallucination, and it increased with slice thickness, though every reconstruction still improved lesion detection over the raw thick slice. ECLARE recovered small lesion signal best at both thicknesses, while the INR was no better than cubic interpolation.
May 13, 2026

On Hallucinations in Inverse Problems: Fundamental Limits and Provable Assessment Methods

Artificial intelligence (AI) has transformed imaging inverse problems, from medical diagnostics to Earth observation. Yet deep neural networks can produce hallucinations, realistic-looking but incorrect details, undermining their reliability, especially when ground truth data is unavailable. We develop a theoretical framework showing that such hallucinations are not merely artifacts of particular models, but can arise from the ill-posed nature of the inverse problem itself. We derive necessary and sufficient conditions for hallucinations, together with computable bounds on their magnitude that depend only on the forward model. Building on this theory, we introduce algorithms to: (1) estimate the minimum hallucination magnitude achievable by any reconstruction model for a given input; (2) assess the faithfulness of reconstructed details by a given reconstruction model. Experiments across three imaging tasks demonstrate that our approach applies broadly, including to modern generative models, and provides a principled way to quantify and evaluate AI hallucinations.
Feb 23, 2026

A Green Learning Approach to LDCT Image Restoration

This work proposes a green learning (GL) approach to restore medical images. Without loss of generality, we use low-dose computed tomography (LDCT) images as examples. LDCT images are susceptible to noise and artifacts, where the imaging process introduces distortion. LDCT image restoration is an important preprocessing step for further medical analysis. Deep learning (DL) methods have been developed to solve this problem. We examine an alternative solution using the Green Learning (GL) methodology. The new restoration method is characterized by mathematical transparency, computational and memory efficiency, and high performance. Experiments show that our GL method offers state-of-the-art restoration performance at a smaller model size and with lower inference complexity.