Off-Axis Compliant RCM Joint with Near-Isotropic Stiffness and Minimal Parasitic Error
Authors: Federico Mariano, Elena De Momi, Giovanni Berselli, Jovana Jovanova, Just L. Herder, Leonardo S. Mattos
Organizations: Biomedical Robotics Lab, Department of Advanced Robotics, Istituto Italiano di Tecnologia, Genoa, Italy. · Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milan, Italy · DIME, Università di Genova, Genoa, Italy · Department of Maritime and Transport Technology, Delft University of Technology, Delft, The Netherlands. · Department of Precision and Microsystem Engineering, Delft University of Technology, Delft, The Netherlands.
This paper presents an off-axis, monolithic compliant Remote Center of Motion (RCM) joint for neuroendoscopic manipulation, combining near-isotropic stiffness with minimal parasitic motion. Based on the Tetra II concept, the end-effector is placed outside the tetrahedral flexure to improve line of sight, facilitate sterilization, and allow rapid tool release. Design proceeds in two stages: mobility panels are sized with a compliance-based isotropy objective, then constraining panels are synthesized through finite-element feasibility exploration to trade stiffness isotropy against RCM drift. The joint is modeled with beam elements and validated via detailed finite-element analyses, including fatigue-bounded stress constraints. A PA12 prototype is fabricated by selective laser sintering and characterized on a benchtop: a 2 N radial load is applied at the end-effector while a 6-DOF electromagnetic sensor records pose. The selected configuration produces a stiffness-ellipse principal axis ratio (PAR) of 1.37 and a parasitic-to-useful rotation ratio (PRR) of 0.63%. Under a 4.5° commanded rotation, the predicted RCM drift remains sub-millimetric (0.015-0.172 mm). Fatigue analysis predicts a usable rotational workspace of 12.1°-34.4° depending on direction. Experiments reproduce the simulated directional stiffness trend with typical deviations of 6-30%, demonstrating a compact, fabrication-ready RCM module for constrained surgical access.
Figures & tables
Fig. 1 : (a) Representation of a double parallelogram RCM mechanism. The mutual parallelism of A, B, C, and D forces the prolongation of the EE to pass through the RCM - (b) Representation of a RCM mechanism with circular guides.
Fig. 2 : View of dual compliant joint composed of two identical joint connected in series. This joint can move the EE around the RCM along the pitch and roll directions. [ 17 ]
Fig. 3 : Isometric view of Tetra II joint composed of nine flexural walls ( p1,...,p9 ). The blue panels, obtained connected the nodes N1,N2,N3 and N4 , are responsible for joint motion and are therefore the ones to be modified to have isotropic stiffness behavior. The red panels block the joint’s rotation, constraining the tool from following circular rotations around the RCM.
Fig. 4 : Scheme of the line of sight of the surgeon over the entry area. The green cone represents the visible area while the black cone represents the non visible area. In (a) the joint has the EE located at the center of tetrahedral structure, while in (b) the joint has the EE located at the side of tetrahedral structure. The EE is represented with the red line.
Fig. 5 : Scheme of the three panels responsible for the joint movement. Highlighted are nodes N1,N2,N3 and N4 , along with the outputs of the optimization L1,L2,L3,t1,t2,t3,θ1 and θ2 .
Parameter
Type
Interval
Lref
Length
20mm≤Lref≤100mm
tref
Thickness
1mm≤tref≤6mm
t△
Thickness
1mm≤t△≤6mm
H
Height
20mm≤H≤100mm
α
Angle
10∘≤α≤70∘
TABLE I : Parameter ranges
Fig. 6 : Representation of the joint from frontal view in (a) , lateral view in (b) and isometric view in (c) . In blue, the structure is responsible for the movement of the end effector, and in red, the constraining structure locks the RCM during the movement. Highlighted are the nodes N1,N2,N3 and N4 with the angle α and all the thickness t1,t2,t3 and t△ .
Fig. 7 : Workflow of the study: MATLAB identifies candidate geometries that satisfy the isotropic-stiffness condition for the mobility panels. A parametric FEM model is then evaluated and the metrics IsoErr and ParErr are computed from the output displacements. A feasibility analysis selects admissible designs and returns Solutions 1–4. These candidates are re-simulated with FEM, and the best configuration is chosen using the Principal-Axis Ratio PAR , the Parasitic-to-useful Rotation Ratio PRR , and the composite index J .
Fig. 8 : Experimental setup used to validate the joint in real-life conditions. A mass producing a force of 2N is attached at point N4, while an Aurora microsensor, operating within the Aurora electromagnetic field, is used to measure the movement of point N4.
Fig. 9 : (a) 3D model of the final compliant joint. (b) Polar graph showing
Parameter
Definition
Value / Ratio
L1
reference length
Lref
L2/L1
—
3.12
L3/L1
—
2.00
t2
reference thickness
tref
t1/t2
—
1.03
t3/t2
—
2.01
TABLE II : Parameters of the isotropic stiffness solution.
Case
Lref [mm]
tref [mm]
PAR
1
20
1
1.12
2
40
1
1.06
3
40
2
1.10
TABLE III : Verification of isotropic stiffness via ANSYS for three geometric scalings.
Conf. 1
Conf. 2
Conf. 3
Conf. 4
tref [mm]
1.59
1.15
1.33
1.83
H [mm]
97.56
84.01
95.98
67.59
Lref [mm]
64.10
70.84
97.86
99.76
t△ [mm]
3.35
3.55
3.42
1.67
α [ ∘ ]
34.40
46.33
42.68
26.61
IsoErr
0.12
0.22
0.27
0.52
TABLE IV : configuration obtained from the feasibility analysis
Conf. 1
Conf. 2
Conf. 3
Conf. 4
PAR
1.17
1.29
1.37
1.87
PRR
0.032
0.015
0.0063
0.0041
J
0.0055
0.0043
0.0023
0.0036
TABLE V : configuration obtained from the feasibility analysis
Fig. 10 : Graphs of two feasibility metrics, IsoErr (left column, in blue) and ParErr (right column, in red) versus the five inputs. ( a-b ) Lref , ( c-d ) H , ( e-f ) tref , ( g-h ) t△ , and ( i-j ) α . In each graph, the shown variable sweeps its range while the other four are fixed at best configuration obtained after the optimization ( Lref=97.86mm , H=95.98mm , tref=1.33 , t△=3.42mm and α=42.68∘ .
θf [ ∘ ]
Stiffness [N/m]
Parasitic error [mm]
0
285.27
0.066
10
301.02
0.092
20
318.18
0.117
30
334.22
0.141
40
345.61
0.159
50
349.08
0.170
TABLE VI : Stiffness and parasitic error for a 4.5 ∘ rotation.
Fig. 11 : In blue, the rotational workspace of the joint around θf . - In red is the minimum workspace for superimposing the EE over the surgical trajectory.
Stiffness [N/m]
Angle [ ∘ ]
Exp. mean ± SD
Sim.
% Error
0
374.05 ± 6.14
285.27
23.73
30
479.45 ± 30.22
334.22
30.29
60
423.41 ± 17.67
343.48
18.88
90
333.21 ± 28.82
297.02
10.86
120
289.33 ± 4.04
261.08
9.76
TABLE VIII : Comparison between experimental and simulated stiffness values, with associated percentage error.
Metric
Value
MAE [N/m]
63.19
RMSE [N/m]
73.92
MAPE [%]
16.88
Bias [N/m]
60.15
TABLE IX : Global error metrics for the comparison between experimental and simulated stiffness values.
Mechanism
Description
Key features
Tetra II [ 15 ]
A compliant RCM joint based on a tetrahedral flexure architecture, which maintains a fixed remote center of motion while steering the EE.
The EE is located on the central axis of the tetrahedral structure, which limits surgical usability in the operating room.
Off-axis compliant mechanism (unoptimized) [ 17 ]
The EE holder is shifted from the center of the tetrahedral structure to a lateral (off-axis) position to improve surgical accessibility.
Stiffness is highly anisotropic ( PAR=21.63 ) and the peak parasitic error is relatively high ( 1.10mm ).
Dual-joint compliant mechanism [ 17 ]
Two identical off-axis joints are connected in series to reduce parasitic RCM displacement and improve stiffness isotropy.
Stiffness anisotropy is reduced (PAR = 1.55) with a lower peak parasitic error (0.051 mm).
Off-axis compliant joint (optimized)
Starting from the Tetra II architecture, the geometry is optimized, via feasibility analysis, to reduce parasitic motion while maintaining near-isotropic stiffness.
H. Tian, A. Namour, L. Zeng, Z. Wu, S. Ourselin and C. Bergeles are with the School of Biomedical Engineering & Imaging Sciences, King’s College London, UK. · L. da Cruz is with Moorfields Eye Hospital, NHS Foundation Trust, London, UK.
Department of Translational Surgical Oncology, NCT/UCC Dresden, a partnership between DKFZ, MF UKD, TUD, HZDR Dresden, Germany · Cluster of Excellence-CeTI, TUD, Germany · Department of Visceral, Thoracic and Vascular Surgery, Faculty of Medicine and University Hospital Carl Gustav Carus, TUD, Germany