Autonomous Quantum Measurement with Agentic AI
Organizations: Quantum Measurement Group, MIT, Cambridge, MA 02139, USA · Department of Nuclear Science and Engineering, MIT, Cambridge, MA 02139, USA · Department of Electrical Engineering and Computer Science, MIT, Cambridge, MA 02139, USA · Center for Computational Science & Engineering, MIT, Cambridge, MA 02139, USA ·
Abstract
Quantum materials exhibit diverse electronic and functional phenomena, but their characterization requires extensive experimental expertise, iterative decision-making and accumulated laboratory knowledge that is difficult to document and transfer. Here we introduce Quantum Agent (QUANTA), an agentic artificial intelligence (AI) framework for autonomous quantum-materials characterization. Using a widely adopted Physical Property Measurement System (PPMS) as an experimental platform, QUANTA combines retrieval augmentation using scientific literature, experimental protocols and prior measurements with specialized agents for experimental planning, troubleshooting, and data interpretation. A separate deterministic control layer validates instrument actions. Guided by incoming data, QUANTA evaluates measurement quality, revises experimental plans and selects subsequent measurements. We demonstrate these capabilities through transport measurements of the topological Weyl semimetal TaP, as well as investigations of nonlinear responses in magnetic material Co2MnGa. These results establish QUANTA as a generalizable framework for evidence-driven, safeguarded and closed-loop quantum-materials characterization.
Figures & tables
Appendix figures & tables12 assets
Supplementary material from the paper’s appendix.
Appendix
| Layer | Required content and role |
| Researcher intake | Objective, sample identity, physical wiring evidence, allowed temperature, field and excitation range, time budget, completion criterion, excluded physical actions and stop authority. Establishes campaign scope but does not encode an executable command |
| Repository entry instructions | Exact deployment, worker owner, configuration, state directory and requirement to read the operating guide. Prevents reuse of a stale endpoint or a different sample’s profile |
| Operating guide | Manual and literature retrieval, sequence search, qualification, validation, monitoring, troubleshooting and shutdown policy. Defines how a scientific objective is converted into bounded stages |
| Campaign record and checklist | Hypotheses, controls, authorization, limits, accepted and rejected evidence, pending route and final-state requirement. Maintains the evolving scientific state and handoff record |
| Plan and deployment configuration | Ordered set points, rates, electrical settings, repetitions, stop conditions, numerical ceilings and configuration hash. Supplies the only representation accepted by the validator and worker |
| SQLite run record | Plans, runs, readings, events, cached state, timestamps, quality flags and shutdown result. Connects a scientific decision to the exact action and returned evidence |
| Case | Evidence, guarded action and interpretation limit |
| TaP overload, 300 K, 0 T | Evidence: At RMS and , the SR860 overloaded with a input range and sensitivity; no reading was accepted. Action: Stop excitation, confirm source-off and retest in a new validated plan with a range and sensitivity. Conclusion: The retest produced 12/12 readable values, establishing recovery but not the physical overload cause |
| TaP current bracket, 100 K, 0 T | Evidence: Within a 1–3–1 mA RMS bracket, the 3-mA run and subsequent 1-mA return differed by in . Their combined descriptive block uncertainty was , giving a three-times-uncertainty threshold of , which exceeded the alternative 1% signal-scale floor. The observed difference exceeded this threshold. Action: Withhold the 3-mA field scan and return to a longer-settled 1-mA plan. Conclusion: The 1-mA scan yielded 180/180 valid readings; heating, non-ohmic response, contact change, pickup and drift were not separated, and 3-mA damage was not established |
| TaP low-current bracket, 2 K, 0 T | Evidence: In the 1–0.5–1 mA RMS comparison, the 0.5-mA run gave , where and are the mean current-normalized components and and are their four-block standard uncertainties. This fell below the selected detectability cutoff of 10. Action: Do not promote 0.5 mA as an equivalent high-field setting. Conclusion: No mean shift was resolved, but equivalence remained unproven because detectability failed |
| TaP timing test, 2 K, +9 T | Evidence: Four 61-reading traces compared 1-s and 3-s time constants. Two traces per setting, three candidate delays for 1 s, two for 3 s and three window sizes produced analysis windows. The selected precision target was , subject to validity, correlation, batch-count, stationarity and repeatability checks. This target was a standard uncertainty, not a 95% interval half-width. Action: Retain all traces and withhold a faster production setting until all acceptance criteria pass. Conclusion: All 30 windows were inconclusive; unresolved repeatability was not treated as zero, and neither a timing setting nor a net time saving was qualified |
| Campaign configuration | Recorded action envelope and interpretation |
| TaP AC transport | Environment: 2–301 K and 9\text{,}\mathrm{T} ; maximum rates $2\text{\,}\mathrm{K}\text{\,}{\mathrm{min}}^{-1}$ and $0.01\text{\,}\mathrm{T}\text{\,}{\mathrm{s}}^{-1}$ . Electrical: $I_{\mathrm{peak}}\leq$4.3\text{\,}\mathrm{mA} , compliance 0.1\text{,}\mathrm{V}$$ and configured estimated power . Run bound: 3,000 points, 50,000 readings and 24 h. The current ceiling was a conservative qualification boundary, not a demonstrated sample-safe limit |
| Co 2 MnGa out-of-plane Hall | Environment: 2–300 K and 9\text{,}\mathrm{T} ; maximum rates $3\text{\,}\mathrm{K}\text{\,}{\mathrm{min}}^{-1}$ and $0.02\text{\,}\mathrm{T}\text{\,}{\mathrm{s}}^{-1}$ . Electrical: $I_{\mathrm{peak}}\leq 0.30\,\mu\mathrm{A}$ , compliance $\leq$0.1\text{\,}\mathrm{V} and configured estimated power 10\text{,}\mathrm{nW}$$ . Run bound: 10,000 points, 100,000 readings and 24 h. Both voltage channels were read sequentially; a transverse channel label did not by itself establish Hall polarity |
| Co 2 MnGa in-plane nonlinear campaign | Environment: 2–300 K and 1\text{,}\mathrm{T} ; maximum rates $4\text{\,}\mathrm{K}\text{\,}{\mathrm{min}}^{-1}$ and $0.01\text{\,}\mathrm{T}\text{\,}{\mathrm{s}}^{-1}$ . Electrical: $I_{\mathrm{peak}}\leq$1.5\text{\,}\mathrm{mA} , compliance 5\text{,}\mathrm{V} and port-power screening $\leq$15\text{\,}\mathrm{mW} ; commissioned transport used at most RMS with a source-on timer. Run bound: 10,000 points, 100,000 readings and 6 h. Port-power screening was not a measurement of sample or contact dissipation |
| Observation class | Response and remaining limitation |
| Stale deployment, owner mismatch or unverified wiring map | Hold dispatch; reconcile worker, configuration and physical record. Unresolved: actual lead continuity, grounding and sample identity require independent evidence |
| Malformed or out-of-envelope plan | Reject before queueing and return the failed field or numerical bound. Scientific desirability cannot override a hard limit |
| Source compliance, overtemperature or inconsistent source state | Stop the affected run, preserve flags and attempt verified source-off. Unresolved: programmed current is not an independent measurement of sample current |
| SR860 input overload or reference loss during excitation | Reject affected readings, stop when required and use a new validated diagnostic plan. Unresolved: a range change can restore acquisition without identifying the physical cause |
| Chamber fault, non-finite or persistently unknown state | Pause or fault, attempt source-off and retain the last PPMS targets for explicit review. Unresolved: chamber mode is not a pressure measurement; electrical shutdown does not zero the magnet |
| Communication or worker fault | Preserve durable run state, block blind restart and reconcile ownership before a new run. Unresolved: process or bus failure can prevent a guaranteed shutdown; this is not an emergency-stop system |
Autonomous Experimentation
Autonomous experimentation uses AI, active learning, and robotics to design, run, and interpret experiments with less human intervention.