When can multi-fragment 2D/3D registration be trusted? Calibrated certification and uncertainty-driven view planning for pelvic fracture fluoroscopy

Authors: Bowen Xiang

Organizations: Department of Biomedical Engineering, Vanderbilt University, 2301 Vanderbilt Place, Nashville, TN 37235, USA

Abstract

Registering a segmented CT to fluoroscopy can locate every fracture fragment but can guide surgery only when trusted. After each shot, our closed loop registers all fragments, bounds their errors with a calibrated sandwich covariance, flags implausible or ambiguous ones, and stops or plans the next view. On simulated fluoroscopy of 60 held-out fracture CTs with realistic image uncertainty, 95% bounds covered 83% of unflagged fragments and 91% of those within 5mm, but from the second view on 14% failed by over 5mm, a third with bounds of at most 2mm. A fragment's shift when a view was added separated tightly bounded failures best (exploratory; AUC 0.94, bound 0.81). Requiring that the newest view moved no fragment beyond its bound, preferred post hoc, lowered false stops at 2mm from 23% to 6% overall, with 345 instead of 362 correct ones but more trajectories uncertified; delaying the stop did not, and other signals at equal acceptance left 10-15% false. With a segmentation error shared by all views (median Dice 0.89), 21% of this rule's stops were false against 10% without it (15 cases; exploratory). Minimax planning certified sooner than the clinical protocol without the check (4.20 versus 4.72 shots, Holm p <0.001), but no criterion reached the pre-specified endpoint, a true 2mm error, significantly sooner, even in the true basin. On real fluoroscopy of five intact pelves no 2mm certificate was false. Trust takes evidence across views, not only a calibrated local bound.

Figures & tables

Appendix figures & tables17 assets

Supplementary material from the paper’s appendix.

Appendix

Explore similar work

Aug 7, 2026

RibAssist 3D: Biplanar Rib-Fracture Detection, Addressing, and Selective 3D Localization from CT-Derived Projections

Rib fractures are common and time-consuming to localize on computed tomography (CT). We ask whether fractures detected independently in two orthogonal CT-derived projections (anteroposterior and lateral) can be paired across views and triangulated into reliable 3D points at a controlled rate of false outputs, and we answer it with a staged diagnostic study. The projection geometry is exact, and given correct correspondence, localization is accurate (median 4.0 mm, 88% within 10 mm, 93.6% rib-exact). On a sealed 55-case cohort, a large share of fractures is in principle recoverable (61.1% dual-view availability, and a correct pair present in the candidate graph for 58.4% of fractures), yet the binding limitation is neither geometry nor localization but confidence-limited cross-view correspondence. A controlled detector-by-correspondence factorial attributes the operational gain to lateral-detector quality rather than the tested matching methods; retraining the lateral detector produces the first nonzero controlled-budget reconstructions. Under a deliberately conservative commitment policy, a pre-specified sealed pass promotes 15 of 601 fractures to correct 3D localizations at 0.436 false points per case (2.50% end-to-end commitment yield), and committed points are accurate (median 1.49 mm, 93% rib-exact). The low yield is a consequence of confidence-gated abstention, not of geometry or detection: the study establishes a reproducible framework for selective 3D localization and identifies cross-view correspondence as the dominant operational bottleneck.
Oct 15, 2026

Bridging the Markerless Tracking Validation Gap: Intraoperative X-ray Workflow Integration

Markerless intraoperative tracking of spinal anatomy via RGB-D sensors is a promising alternative to invasive bone-anchored fiducials, yet no existing clinical dataset provides independent subsurface ground truth for validating these methods. We address this gap with a clinically viable pipeline that derives per-vertebra 6-DoF pose from routine intraoperative C-arm fluoroscopy, co-registered to the RGB-D camera frame through a multimodal fiducial array detectable in both imaging modalities. Critically, the pipeline requires no additional radiation, instrumentation, or deviation from standard surgical workflows, relying solely on the fluoroscopic images already acquired during routine level checks. A calibrated transform chain links the vertebral pose recovered via multi-view 2D-3D CT-to-fluoroscopy registration to the RGB-D coordinate frame, enabling direct comparison of any markerless registration estimate against a subsurface reference. Ex vivo validation against optically tracked ovine vertebrae confirms end-to-end accuracy of 2.06 +- 0.29 mm and 3.35 +- 0.93 degrees, sufficient to serve as independent ground truth for methods with substantially larger errors. We demonstrate the pipeline's utility by evaluating two markerless registration strategies on clinical data spanning 11 vertebrae across 1,500 frames from three surgical sequences: the first such evaluation against an independent subsurface reference in a clinical setting. The proposed framework provides reusable infrastructure for benchmarking emerging markerless tracking methods on real surgical data.
Oct 7, 2026

ActiveReg: Information-Driven Active Regional Probing for Partial-to-Full Bone Registration

Accurate bone registration is essential for orthopedic surgical navigation and robotic assistance. In a common workflow, the surgeon needs to identify and probe a number of prescribed locations on the exposed bone surface to obtain data points for registration, which can be difficult and time-consuming under limited surgical exposure. Reducing the number of required points while providing clear probing guidance would ease the surgeon's acquisition task intra-operatively. We present ActiveReg, a closed-loop framework that recommends probing regions rather than individual points, allowing flexibility in the exact contact location. A D-optimal planner that takes into account the already acquired points and the uncertainty of current registration information helps achieve accurate registration with fewer points. An online assessment combines pre-update innovation consistency with uncertainty at predefined surgical task locations improves the reliability of completion decision, without requiring ground truth. Simulations across four anatomical scenarios demonstrated that ActiveReg can achieve competitive accuracy compared with baseline registration methods while using substantially fewer acquired points. Real phantom experiments with optical and electromagnetic tracking demonstrated comparable mean target registration error to the Gradient-SDF framework, using only 24-28 acquired points instead of 771-1263.