MRI-Driven Dual-Energy CT Synthesis via Conditional Denoising Diffusion Probabilistic Models for MRI-only Radiotherapy and Protontherapy

Authors: Juan Manuel Molina-Maza Chih-Wei Chang, David Viar-Hernandez, Zach Eidex, Mojtaba Safari, Shaoyan Pan, Juan Maria Perez-Moreno, Norberto Malpica, Angel Torrado-Carvajal, 

Organizations: Department of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, USA · Medical Image Analysis and Biometry Laboratory, Universidad Rey Juan Carlos, Madrid, Spain

Abstract

Magnetic resonance imaging (MRI) provides excellent soft-tissue contrast without ionizing radiation but lacks the electron-density information needed for radiotherapy dose calculation. Dual-energy computed tomography (DECT) improves tissue characterization and stopping-power ratio (SPR) estimation for proton therapy, but requires additional radiation exposure. We developed a denoising diffusion probabilistic model (DDPM) to synthesize high- and low-energy CT (sHECT and sLECT) simultaneously from a single MRI scan. Retrospective paired MRI-DECT data from patients receiving head-and-neck proton therapy were used to train and evaluate a multi-decoder DDPM with SwinViT or ResUNet denoisers. Performance was compared with Pix2PixGAN and CycleGAN using image-quality metrics, Hounsfield unit (HU) errors, anatomical profiles, and SPR maps and uncertainties. The DDPM outperformed both generative adversarial network baselines in visual fidelity and quantitative image quality. For sHECT and sLECT, respectively, mean absolute errors were 36.29 +/- 5.25 and 43.72 +/- 6.77 HU; peak signal-to-noise ratios were 27.44 +/- 1.06 and 26.26 +/- 0.98 dB; structural similarity indices were 0.89 +/- 0.02 and 0.87 +/- 0.03; and normalized cross-correlations were 0.95 +/- 0.01 and 0.94 +/- 0.01. The synthesized images preserved anatomical boundaries, while uncertainties in derived SPR maps remained within the expected acceptable range. MRI-based DECT synthesis using DDPM may provide tissue-specific information for proton therapy planning while reducing the need for radiation-based DECT acquisition.

Figures & tables

Explore similar work

May 13, 2026

Generating synthetic computed tomography for radiotherapy: SynthRAD2025 challenge report

Radiation therapy (RT) requires precise dose delivery over multiple fractions, with CT fundamental for treatment planning due to its electron density information. Repeated CT acquisitions impose radiation exposure and logistical burdens, MRI lacks electron density, and cone-beam CT (CBCT) requires correction for dose calculation. Synthetic CT (sCT) generation addresses these by converting MRI or CBCT into CT-equivalent images with accurate Hounsfield Unit (HU) values, enabling MRI-only RT and CBCT-based adaptive workflows. Building on SynthRAD2023, SynthRAD2025 benchmarked sCT methods on 2,362 patients from five European centers across head and neck, thorax, and abdomen. Two tasks: MRI-to-CT (890 cases) and CBCT-to-CT (1,472 cases), evaluated via image similarity (MAE, PSNR, MS-SSIM), segmentation (Dice, HD95), and dosimetric metrics from photon and proton plans. With 803 participants and 12/13 valid submissions, Task 1 top performance reached MAE 64.8±21.364.8\pm21.3 HU, PSNR ∼\sim30 dB, MS-SSIM ∼\sim0.936, Dice 0.79, photon γ2%/2mm>98%γ_{2\%/2\text{mm}}>98\%, proton γ≈85%γ\approx85\%. Task 2 improved: MAE 48.3±13.448.3\pm13.4 HU, PSNR 32.6 dB, MS-SSIM 0.968, Dice 0.86, photon γ>99%γ>99\%, proton γ≈89%γ\approx89\%. Strong image--segmentation correlations (ρ=0.78ρ=0.78--0.790.79) but moderate dose correlations confirmed image quality is insufficient as a dosimetric surrogate. Head-and-neck cases were most consistent; thoracic and abdominal cases showed greater variability. Residual errors at tissue interfaces propagate along beam paths, affecting proton dose more than photon. SynthRAD2025 demonstrates that deep learning yields clinically relevant sCTs, especially for CBCT-to-CT, while identifying persistent MRI-to-CT challenges and underscoring dose-based evaluation as essential for clinical validation.
Jun 4, 2025

Generalizable dose denoising using diffusion transformer framework for pencil beam scanning proton therapy

Background and purpose: Intensity-modulated proton therapy (IMPT) enables conformal dose delivery for head and neck (H&N) cancers. However, its sensitivity to inter-fractional anatomical changes motivates online adaptations, where rapid and accurate Monte Carlo (MC) dose calculation remains a major bottleneck. Low-statistics MC simulations reduce computation time but introduce noise. This study aimed to develop a diffusion-transformer framework to recover high-statistics dose distributions from noisy MC simulations. Materials and methods: A diffusion-transformer-based dose denoising framework was developed. Eighty H&N patients were used for training and additional 10 H&N patients and 30 patients with lung, breast, and prostate cancers (10 per site) were used for evaluation. Noisy and high-statistics dose maps were generated using MCsquare. The model was trained using noisy dose maps and CT images as inputs and high-statistics dose maps as ground truth, with mean squared error, residual, regional mean absolute error (MAE), and dose-volume histogram (DVH) losses. Performance was evaluated using MAE, 3D gamma passing rates, and DVH indices. Results: Mean whole-body MAE ranged from 0.08 +/- 0.06 to 0.20 +/- 0.11 Gy (Relative Biological Effectiveness, RBE) (RBE = 1.1) across test cohorts. Mean whole-body 3 %/2 mm gamma passing rates ranged from 95.8 +/- 0.8 % to 97.6 +/- 0.9 %. DVH indices from high-statistics and predicted dose distributions showed close agreement for clinical target volumes and organs of interest. Conclusions: The diffusion-transformer-based framework denoised low-statistics MC dose distributions. Despite being trained on H&N data, the model generalized to lung, breast, and prostate cohorts without retraining.
Mar 30, 2026

MRI-to-CT synthesis using drifting models

Accurate MRI-to-CT synthesis could enable MR-only pelvic workflows by providing CT-like images with bone details while avoiding additional ionizing radiation. In this work, we investigate recently proposed drifting models for synthesizing pelvis CT images from MRI and benchmark them against convolutional neural networks (UNet, VAE), a generative adversarial network (WGAN-GP), a physics-inspired probabilistic model (PPFM), and diffusion-based methods (FastDDPM, DDIM, DDPM). Experiments are performed on two complementary datasets: Gold Atlas Male Pelvis and the SynthRAD2023 pelvis subset. Image fidelity and structural consistency are evaluated with SSIM, PSNR, and RMSE, complemented by qualitative assessment of anatomically critical regions such as cortical bone and pelvic soft-tissue interfaces. Across both datasets, the proposed drifting model achieves high SSIM and PSNR and low RMSE, surpassing strong diffusion baselines and conventional CNN-, VAE-, GAN-, and PPFM-based methods. Visual inspection shows sharper cortical bone edges, improved depiction of sacral and femoral head geometry, and reduced artifacts or over-smoothing, particularly at bone-air-soft tissue boundaries. Moreover, the drifting model attains these gains with one-step inference and inference times on the order of milliseconds, yielding a more favorable accuracy-efficiency trade-off than iterative diffusion sampling while remaining competitive in image quality. These findings suggest that drifting models are a promising direction for fast, high-quality pelvic synthetic CT generation from MRI and warrant further investigation for downstream applications such as MRI-only radiotherapy planning and PET/MR attenuation correction.