Intrinsic Tolerance in C-Arm Imaging: How Extrinsic Re-optimization Preserves 3D Reconstruction Accuracy

Authors: Lin Li, Benjamin Aubert, Paul Kemper, Aric Plumley

Organizations: AIX, Alphatec Spine, 1950 Camino Vida Roble, Carlsbad, CA 92008, USA · EOS, EOS Imaging Inc., 7275 Rue Saint-Urbain Suite 400, Montreal, QC H2R 2Y5, Canada

Abstract

Purpose: C-arm fluoroscopy's 3D reconstruction relies on accurate intrinsic calibration, which is often challenging in clinical practice. This study ensures high-precision reconstruction accuracy by re-optimizing the extrinsic parameters to compensate for intrinsic calibration errors. \noindentMethods: We conducted both simulation and real-world experiments using five commercial C-arm systems. Intrinsic parameters were perturbed in controlled increments. Focal length was increased by 100 to 700 pixels (≈\approx20 mm to 140 mm) and principal point by 20 to 200 pixels. For each perturbation, we (1) reconstructed 3D points from known phantom geometries, (2) re-estimated extrinsic poses using standard optimization, and (3) measured reconstruction and reprojection errors relative to ground truth. \noindentResults: Even with focal length errors up to 500 pixels (≈\approx100 mm, assuming a nominal focal length of ∼\sim1000 mm), mean 3D reconstruction error remained under 0.2 mm. Larger focal length deviations (700 pixels) elevated error to only ≈\approx0.3 mm. Principal point shifts up to 200 pixels introduced negligible reconstruction error once extrinsic parameters were re-optimized, with reprojection error increases below 0.5 pixels. \noindentConclusion: Moderate errors in intrinsic calibration can be effectively mitigated by extrinsic re-optimization, preserving submillimeter 3D reconstruction accuracy. This intrinsic tolerance suggests a practical pathway to relax calibration precision requirements, thereby simplifying C-arm system setup and reducing clinical workflow burden without compromising performance.

Figures & tables

Explore similar work

CardsList
  1. Beyond Reprojection Error: Camera Calibration with 3D Targets

    Aug 5, 2026Dennis Ruppel, Hasan Kutlu, Kai A. Neumann +43D ReconstructionCamera Calibration

  2. Projection-Domain Sensitivity Analysis of Vertebral DRRs Under Intrinsic Calibration Perturbation

    Jul 12, 2026Lin Li, Chaochao Zhou, Benjamin Aubert +2Medical ImagingMedical Image Registration

  3. Automatic Reproducible Camera Intrinsic Calibration

    Sep 9, 2026Xiangcheng HuCamera Calibration