AEGIS: An Operational Infrastructure for Post-Market Governance of Adaptive Medical AI Under US and EU Regulations
Authors: Fardin Afdideh, Mehdi Astaraki, Fernando Seoane, Farhad Abtahi
Organizations: Department of Clinical Science, Intervention and Technology, Karolinska Institutet, 17177 Stockholm, Sweden · Department of Medical Radiation Physics, Stockholm University, Solna, Sweden · Department of Oncology-Pathology, Karolinska Institutet, Solna, Sweden · Department of Clinical Physiology, Karolinska University Hospital, 17176 Stockholm, Sweden · Department of Textile Technology, University of Bor˚as, Bor˚as 50190, Sweden · Department of Medical Technologies, Karolinska University Hospital, 141 57 Huddinge, Sweden · Department of Biomedical Engineering and Health System, KTH Royal Institute of Technology, 14157 Huddinge, Sweden
Machine learning systems deployed in medical devices require governance frameworks that ensure safety while enabling continuous improvement. Regulatory bodies including the FDA and European Union have introduced mechanisms such as the Predetermined Change Control Plan (PCCP) and Post-Market Surveillance (PMS) to manage iterative model updates without repeated submissions. This paper presents AI/ML Evaluation and Governance Infrastructure for Safety (AEGIS), a governance framework applicable to any healthcare AI system. AEGIS comprises three modules, i.e., dataset assimilation and retraining, model monitoring, and conditional decision, that operationalize FDA PCCP and EU AI Act Article 43(4) provisions. We implement a four-category deployment decision taxonomy (APPROVE, CONDITIONAL APPROVAL, CLINICAL REVIEW, REJECT) with an independent PMS ALARM signal, enabling detection of the critical state in which no deployable model exists while the released model is simultaneously at risk. To illustrate how AEGIS can be instantiated across heterogeneous clinical contexts, we provide two examples: sepsis prediction from electronic health records and brain tumor segmentation from medical imaging. Both cases use identical governance architecture, differing only in configuration. Across 11 simulated iterations on the sepsis example, AEGIS yielded 8 APPROVE, 1 CONDITIONAL APPROVAL, 1 CLINICAL REVIEW, and 1 REJECT decision, exercising all four categories. ALARM signals were co-issued at iterations 8 and 10, including the critical state where no deployable model exists and the released model is simultaneously failing. AEGIS detected drift before observable performance degradation. These results demonstrate that AEGIS translates regulatory change-control concepts into executable governance procedures, supporting safe continuous learning for adaptive medical AI across diverse clinical applications.
Figures & tables
Dimension
FDA (510(k)/ PCCP )
EU MDR
Core Philosophy
SE
Continuous risk management
Foundational Question
“Is this device as safe and effective as something already proven?”
“Does the ongoing risk-benefit analysis remain acceptable?”
Modifications within PCCP bounds maintain equivalence
Changes assessed against evolving risk profile
Methodological Backbone
Predicate device comparison
ISO 14971:2019 risk management
TABLE I: Regulatory Philosophy Comparison: FDA vs. EU MDR
Fig. 1: DARM architecture. The module manages three datasets: Training ( DT,k ) for parameter updates, Golden ( DG ) as static benchmark, and Drifting ( DD,k ) for drift monitoring.
Parameter Categories
Description
Tolerance
Acceptable deviation from target value
Range
Minimum and maximum bounds for acceptable values
Reference
Reference values against which comparisons are made
Distribution Shift Metrics
KL divergence, JS distance, mean drift
Trustworthy AI (TAI) Metrics
Demographic parity, equalized odds metrics
Model Performance Metrics
Accuracy, sensitivity, specificity, AUC, F1-score
TABLE II: Parameters
Fig. 2: MMM architecture showing metric calculation across performance, fairness, and distribution categories, historical KPI maintenance, and outputs for monitoring and compliance.
Decision
Question answered
Boolean Function f
Operator & Rationale
CHANGE-CONTROL DEPLOYMENT DECISIONS: new candidate model
REJECT
Is new model safe?
(PG,k<Pfail)∨(any safety metric<its floor)
OR — any single safety failure blocks deployment
CLINICAL REVIEW
Does new model need human validation?
(Pfail≤PG,k<Pfail+τ)∨(τ<Pref−PG,k)
OR — any metric in buffer zone, or any regression from Pref (fixed performance reference comparison fails).
CONDITIONAL APPROVAL
Does the new model need enhanced monitoring?
(SD∈SDminor)∨(ST<STthr)
OR — minor drift or any TAI concern each independently warrant enhanced monitoring
AND — all checks must pass; any single failure routes to a lower category
PMS ALARM SIGNAL: currently released model
TABLE III: CDM Decision Categories. Deployment decisions: Four decision categories for a new-model deployment. PMS ALARM signal: ALARM for the currently released model. The two parts are evaluated independently and produce a composite output.
Priority
Check
Condition
Decision
New Model Deployment Hierarchy (evaluate in priority order on DG and DD,k )
P1
deployment safety performance floor
PG,k<Pfail
REJECT
P2
performance buffer / regression from fixed performance reference
(PG,k∈RG)∨(τ<Pref−PG,k)
CLINICAL REVIEW
P3
minor drift / TAI
(SD∈SDminor)∨(ST<STthr)
CONDITIONAL APPROVAL
P4
all pass
all checks passed
APPROVE
PMS ALARM Signal (evaluate in parallel using released model on DD,k )
Fig. 5: AEGIS instantiation pattern. The abstract infrastructure (middle) defines domain-agnostic modules. Each instantiation (top and bottom) configures with domain-specific parameters while maintaining identical logic.
Decision
Definition
FDA PCCP Status
EU MDR / AI Act
Trigger
APPROVE
All metrics satisfied
Within PCCP scope
Art. 43(4) exemption
All pass
CONDITIONAL APPROVAL
Enhanced monitoring
Limited PCCP scope
Enhanced PMS
Minor drift or low trust scores
CLINICAL REVIEW
Human validation required
May exceed PCCP
Pre-modification review
Near threshold
ALARM
Released model at risk in field; issued as a PMS signal independently of the new model deployment decision
Activates PMS reporting; may require vigilance report depending on severity
PMS safety performance floor breach; released model regression; major distributional shift
REJECT
Deployment blocked
New 510(k) required
New CA
Safety violation
TABLE VI: Decision Taxonomy with Regulatory Mapping ∗
AEGIS Parameter
Concrete Value
Clinical Justification
CDM Decision
Pref
Sens.: 0.723, Spec.: 0.933, AUC: 0.922
performance of the first fully approved model on the Golden dataset
fixed performance reference
Pfail
Sens.: 0.65
derived from published systematic reviews of sepsis prediction performance [ 8 , 30 ]
REJECT
RG
0.66≤ Sens. 0.60≤ Spec.
Performance thresholds (buffer zone)
CLINICAL REVIEW
Major drift threshold
0.90<SD
>90% of 34 key features shifted ( KS test)
ALARM (A3)
Minor drift range
0.30≤SD≤0.70
Moderate distributional change detected
CONDITIONAL APPROVAL
PPMS∗
Sens.: 0.65
=Pfail
ALARM (A1)
TABLE VII: Mapping of Abstract AEGIS Parameters to Sepsis Prediction Parameters. This table provides the sepsis-specific instantiation of the infrastructure-level CDM priority hierarchy defined in Table IV .
Fig. 6: GLI tumors in pre-operative and post-treatment mpMRI .
Iter.
∣DT,k∣
Sens.
Spec.
AUC
MLcps
SD
Trigger
Routing Trace ∗
Deploy. Decision
PMS Signal
0
8,134
0.723
0.933
0.922
0.721
0.000
All satisfied
P1 ✓ P2 ✓ P3 ✓→ P4
APPR.
—
1
9,760
0.745
0.926
0.922
0.731
0.000
All satisfied
P1 ✓ P2 ✓ P3 ✓→ P4
APPR.
—
2
11,386
0.745
0.926
0.923
0.731
0.000
All satisfied
P1 ✓ P2 ✓ P3 ✓→ P4
APPR.
—
3
13,012
0.742
0.926
0.923
0.729
0.000
All satisfied
P1 ✓ P2 ✓ P3 ✓→ P4
APPR.
—
4
14,638
0.742
0.932
0.923
0.733
0.000
All satisfied
P1 ✓ P2 ✓ P3 ✓→ P4
APPR.
—
5
16,268
0.753
0.929
0.924
0.739
0.000
All satisfied
P1 ✓ P2 ✓ P3 ✓→ P4
APPR.
—
TABLE VIII: Iteration Summary with Trigger Conditions. The Routing Trace column traces the exact priority-ordered evaluation for the new candidate model at each iteration. The PMS Signal column records ALARM co-issuances where the released model was assessed as at risk.
Fig. 7: AEGIS comprehensive iterative change control results. A) Clinical performance metrics; B) PPV and NPV ; C) ROC-AUC , PR-AUC , and MLcps ; D) Matthews correlation and Kappa; E) Drift score; F) Training size; G) Confusion matrix components; H) Deployment decisions with PMS ALARM markers at iterations 8 and 10; I) MLcps score with CDM decision context.
Iter.
∣DT,k∣
PG,k
PD,k
Trigger
Routing Trace*
Deploy. Decision
PMS Signal
0
1150
0.726 ± 0.313
—
fixed performance reference
—
—
—
1
200
0.646 ± 0.337
0.694 ± 0.325
safety violation ( PG,1<0.676 )
→P1
REJ.
—
2
400
0.673 ± 0.337
0.748 ± 0.289
safety violation ( PG,2<0.676 )
→P1
REJ.
—
3
600
0.690 ± 0.326
0.745 ± 0.308
minor drift ( PGrel=0.690 ; PD,3<PD,2 )
P1✓→P4
APPR.
ALARM
4
800
0.700 ± 0.321
0.758 ± 0.284
All satisfied ( PGrel=0.700 )
P1✓→P4
APPR.
—
5
1000
0.725 ± 0.325
0.735 ± 0.325
minor drift ( PGrel=0.725 ; PD,5<PD,4 )
P1✓→P4
APPR.
ALARM
TABLE IX: DSC Metric for Tumor Core Segmentation with Routing Trace
AEGIS Parameter
Concrete Value
Clinical Justification
CDM Decision
Pref
DSC : 0.726
fixed performance reference. performance of the model trained on the full cohort on the Golden set ( TC region)
Division of Imaging, Diagnostics, and Software Reliability (DIDSR), Office of Science and Engineering Laboratories (OSEL), Center for Devices and Radiological Health (CDRH), U.S. Food and Drug Administration, 10903 New Hampshire Ave, Silver Spring, MD US 20993
Chief Operating Officer and Head of Research and Development, Instil-it, Hyderabad, Telangana. · Chief AI Architect Engineer, Instil-it, Hyderabad, Telangana · Chief Executive Officer, Managing Director and Chief Data Scientist, Instil-it, Hyderabad, Telangana