Repeatability Characterisation and Error Budget of a Consumer Structured-Light Scanner for 3-D Wound Geometry:A Rigid-Phantom Study

Authors: Pushkal KumarAadit AggarwalKarlen Aleksanyan

Abstract

We characterise the measurement error of a free- hand consumer structured-light scanner used to derive three- dimensional geometric wound descriptors. Rigid wound-care phantoms cannot change, so every difference between repeat scans of one site is measurement error; all repeats come from a single scanner unit, nine sites and 23 scans, so this charac- terises one instrument. The 95 percent repeatability limit for reconstructed surface area is a factor of 4.8, with a confidence interval from 3.0 to 7.0, so rescanning an unchanged site can shift the reading from a 79 percent decrease to a 377 percent increase. Forty-four of 45 descriptors fall below an intraclass correlation of 0.50, and none of 248 descriptor and pipeline combinations reaches 0.75. An error budget formed by holding the analy- sis region fixed, leaving sensor and reconstruction untouched, bounds the share of variance attributable to how much surface the operator captured at 75 percent for surface area, 91 percent for hull area and 95 percent for bounding-box diagonal; only hull volume is majority instrumental, at 48 percent, so a better sensor would buy little. No wound is delineated anywhere in the chain, so the comparison against the four-week area reduction used clinically to predict healing, a ratio near 2.1, is a lower bound on the noise an unsegmented pipeline must overcome, not a measurement of wound-area reproducibility. Standardising the analysis region cuts the limit to 2.13, meeting that ratio rather than clearing it. Statistical outlier removal imposes a measured systematic area deficit near 11 percent.

Explore similar work

Aug 13, 2026cs.CV

Validation of Smartphone-Based Photogrammetric 3D Body Scanning for Automated Anthropometric Measurements Compared with a Commercial Depth-Sensor-Based Body Scanner

3D body scanning has become an important tool in healthcare applications because of its rapid and non-invasive nature. While smartphone-based photogrammetric reconstruction provide a low-cost and accessible alternative to commercial 3D body scanners, their performance for whole-body scanning remains insufficiently validated. Thus, we designed this study to comprehensively validate the photogrammetric 3D scanning application by evaluating automatically extracted whole-body measurements and longitudinal body-shape monitoring. We evaluated a representative application, PolyCam, against the commercial depth-sensor-based Fit3D ProScanner using 144 pregnant participants scanned longitudinally throughout pregnancy. We designed an automatic circumference extraction pipeline to get measurements at four anatomical landmarks from paired 3D scans. A linear mixed-effects model was used to evaluate scanner effects and longitudinal body-shape changes. Measurement consistency was assessed using repeated PolyCam scans and tape measurements on a rigid mannequin. PolyCam demonstrated strong agreement with Fit3D, with average biases below 16 mm, intraclass correlation coefficients above 0.8, and Pearson correlation coefficients above 0.9 across all landmarks. Both systems captured comparable longitudinal body-shape changes. Mannequin experiments showed mean biases below 3.5 mm and no significant differences from tape measurements. These findings support smartphone photogrammetry as a potential accessible alternative to commercial body scanners and applicable for longitudinal 3D body-shape assessment.
Ruting Cheng, Boyuan Feng, Chuhui Qiu +4
Date pendingcs.CV

A Calibration Audit of Confidence in Feed-Forward 3D Reconstruction Models

Feed-forward 3D reconstruction models output a per-pixel confidence that is used by downstream systems as an uncertainty signal. The confidence is trained to serve as a weight in the training loss of models. Whether the confidence can be used as an uncertainty magnitude has not been measured. We audit seven backbones on 13 datasets and score the confidence on four properties, i.e., ranking of error, ratio of error to uncertainty on average, slope of this ratio across the confidence range, and coverage of the implied error distribution. Although the confidence ranks error quite well, the uncertainty decoded from the confidence is too small compared to the actual error. The uncertainty has the right size only under the exact training conditions. The median case is off by at least 2.4x across all seven models, while the uncertainty is further off the more confident the model is. Our work shows that the overconfidence appears on unseen scenes even when the model reaches its loss's optimum. As a post-hoc repair we fit a power law on the confidence with two constants per backbone--dataset pair. The repair brings all four audited properties to target at the dataset level, while leaving ranking untouched. Fitted with the target dataset held out, the constants bring the median case from 2.4x off to 1.35x. The repair does not hold below the dataset level, where two-thirds of held-out scenes are still more than five points off in coverage. We attribute what the repair cannot reach to the model, which carries neither the scale of the error nor the shape of its distribution across predictions. We release the audit protocol, its results, and the fitted constants per backbone-dataset pair.
Nanxing Nick Deng, Qing Cheng, Niclas Zeller +1
Jul 25, 2026cs.CV

Hold-Out Self-Validation Cannot Certify Photogrammetric Accuracy: Saturation and Blindness to Coherent Distortion

Internal self-consistency cannot certify the accuracy of a photogrammetric reconstruction, and the failure is structural rather than a matter of tuning. This matters because hold-out self-validation scores are increasingly offered as quality evidence for metric deliverables whose correctness is otherwise unknown without an external survey. We formalise a track-leakage-free hold-out protocol: a deterministic image subset is withheld, and each withheld view is re-localised against only those 3D points supported by two or more retained images, so no view is tested against structure it helped create. We evaluate it on five GNSS-referenced captures across four sites, 13 ETH3D scenes, a EuRoC flight and 30 IMC 2025 scenes. The protocol is well-posed but does not measure accuracy. It saturates: the internal confidence score stays pinned at 1.00 while true error swings 14.1x within one capture. It is blind to coherent distortion: fragmenting corruption is caught, but internally self-consistent, globally distorted models are not, and were wrong by 55-106 m at confidence 1.00 at three of four captures. On IMC 2025 it separates failed from successful reconstructions (rho = 0.68) yet ranks nothing among the successful (rho = 0.01). Track-leakage-free hold-out measures internal geometric consistency: a fragmentation warning, not a substitute for control-point accuracy assessment.
Behnam Asadi