cs.CVOct 6, 2026

LARK: A Low-Cost, Accurate, Occlusion-Resilient, Kalman Filter-Assisted Tracking System for Image-Guided Surgery

Authors: George Sideris, Justin Cree, Andrew Stirling, Mamadou Ly, Étienne Léger, D. Louis Collins

Organizations: Department of Mechanical Engineering, McGill University, Montreal, Canada · Montreal Neurological Institute, McGill University, Montreal, Canada

Abstract

Image-guided surgery (IGS) depends on accurate tracking of surgical instruments to provide real-time navigation relative to anatomical structures. Commercial stereo infrared trackers are accurate but prone to occlusion and cost-prohibitive for many settings. This work presents LARK, a multi-camera optical tracking system using commodity RGB hardware and multi-view redundancy and fusion. We develop and evaluate two complete tracking methods: multi-view monocular pose fusion and multi-view triangulation. Both methods are assessed under varying occlusion levels using a precision-machined grid and an anatomical head phantom, and compared against a gold-standard stereo infrared system. With five cameras and adaptive Kalman filtering, LARK achieves median target registration errors of 0.64 mm for point localization with triangulation and 0.73 mm for trajectory tracking with pose fusion on the machined grid. Camera-subset experiments show graceful degradation in adaptive pose-fusion accuracy as fewer views remain available. With tracking hardware costing under $1,000 USD, LARK provides a low-cost platform for image-guided surgery research. Hardware designs and software are publicly available at https://nist.mni.mcgill.ca/software/ , and datasets at https://nist.mni.mcgill.ca/data/ .

Figures & tables

Appendix figures & tables6 assets

Supplementary material from the paper’s appendix.

Appendix

Explore similar work

Sep 17, 2026cs.RO

LapaTrack-3D: 6 DoF pre-operative shape tracking for laparoscopic surgery

This work proposes a real-time 6 Degree-of-Freedom (6 DoF) tracking algorithm for monocular laparoscopic surgery. It provides alignment between intra-operative video and pre-operative data (e.g., CT). The 6 DoF tracking offers a solution for accurately locating the internal anatomy of the target organ despite the lack of tactile feedback and transparency. The ORB-SLAM2 framework is adopted and modified for prior-based 3D tracking with four major modifications. First, the primitive 3D shape is used for fast initialization of the ORB-SLAM2 monocular mode. Second, a pseudo-segmentation strategy is employed to separate the target organ from the background for tracking. Third, the 3D shape is incorporated as a geometric prior in its pose graph optimization. Fourth, the Multi-Scale Retinex with Chromaticity Preservation (MSRCP) algorithm is leveraged and modified for image enhancement in challenging illumination scenarios. In-vivo and ex-vivo experiments validate that LapaTrack-3D provides robust 3D tracking and effectively handles typical challenges such as poor illumination, fast motion, out-of-field-of-view scenarios, partial visibility, and ``organ-background'' relative motion. LapaTrack-3D achieves a processing rate of 13 Hz for 1280*720 pixel video.
Sep 22, 2026cs.RO

Recording Hand-Held Laparoscopic Instrument Motion in the Operating Room: Magnetometer-Free Fusion of Inertial, Range and Visual Sensing

Most minimally invasive procedures are still performed with hand-held laparoscopic instruments, yet only the endoscopic video is retained; the instrument motion that expresses surgical skill, and that could support skill assessment and robot learning, is lost. Pose from video alone remains millimeters to centimeters off, and an instrument-mounted inertial measurement unit (IMU) cannot rely on its magnetometer, whose field changed with tool pose and between sessions in our measurements. We present a surgical instrument-state logger that clips onto a conventional instrument without modifying the part that enters the patient and fuses a six-axis IMU and a time-of-flight (ToF) rangefinder with a markerless camera in an error-state Kalman filter under the remote center of motion (RCM) of the trocar. Heading comes from the shaft silhouette, segmented by a U-Net, in place of the magnetometer: the rotation-angle error is 0.200°, against 3.58° from the accelerometer and magnetometer alone. Against a Franka Research 3 manipulator, and without alignment to it, the displacement error over 300 translation trials was 1.21mm RMS and the relative-rotation error over 180 rotation trials 0.34° RMS. On continuous trajectories, tracked and displayed in real time, the absolute tip error was 1.22mm (programmed) and 3.04mm (teleoperated) after post-hoc tuning of three filter parameters, and the full fusion beat every sensor subset. Because the estimator uses no magnetic measurement, its accuracy does not rely on an undisturbed field. The same clip-on device could thus record metric tip trajectories during routine hand-held laparoscopy, while displaying the insertion depth and attitude that are hidden once the instrument is inside the patient.
Mar 12, 2026cs.RO

Real-time Rendering-based Surgical Instrument Tracking via Evolutionary Optimization

Accurate and efficient tracking of surgical instruments is fundamental for Robot-Assisted Minimally Invasive Surgery. Although vision-based robot pose estimation has enabled markerless calibration without tedious physical setups, reliable tool tracking for surgical robots still remains challenging due to partial visibility and specialized articulation design of surgical instruments. Previous works in the field are usually prone to unreliable feature detections under degraded visual quality and data scarcity, whereas rendering-based methods often struggle with computational costs and suboptimal convergence. In this work, we incorporate CMA-ES, an evolutionary optimization strategy, into a versatile tracking pipeline that jointly estimates surgical instrument pose and joint configurations. Using batch rendering to efficiently evaluate multiple pose candidates in parallel, the method significantly reduces inference time and improves convergence robustness. The proposed framework further generalizes to joint angle-free and bi-manual tracking settings, making it suitable for both vision feedback control and online surgery video calibration. Extensive experiments on synthetic and real-world datasets demonstrate that the proposed method significantly outperforms prior approaches in both accuracy and runtime. Source code and data are available at https://github.com/hanyang-hu/online_dvrk_tracking.