Integrated Respiratory Sensing and Gated Control for Autonomous Needle Insertion Via CT Bore Robotic Arm
Authors: Emma Huang, Moira Bohley, Krishna Jamakhandi, Rebecca Fellbaum, Sophia Tang, Sean Tutton, Albert Hsiao, Alexander Norbash, Changwei Chen, Derek Chen, Ananya Rao, Lizzie Peiros, Michael C. Yip
Organizations: Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, CA 92093 USA. · Department of Bioengineering, University of California San Diego, La Jolla, CA 92093 USA. · Columbia Engineering, Columbia University, New York, NY 10027 USA. · School of Medicine, University of California San Diego, La Jolla, CA 92093 USA. · School of Medicine, University of Missouri-Kansas City, Kansas City, MO 64110 USA.
Currently, CT-guided lung biopsy is a highly manual procedure requiring patient breath-holding, iterative manual needle insertions, and repeated CT confirmation scans to accurately target a tumor. Automating this process within the CT bore--without relying on manual needle placement or patient breath-holding--has the potential to reduce procedure time, patient discomfort, and radiation exposure. In this work, we present an integrated sensing and control system capable of performing a gated needle insertion within the CT bore. The experiments demonstrate the system's ability to track and predict patient breath cycles, perform autonomous respiratory-gated needle insertion, and detect breath cycle anomalies.
Figures & tables
Parameter
Value
Observed range
Literature reference values
Breath duration (s)
4.335
–
Displacement amplitude (mm)
4.33
–
Inhalation/exhalation ratio
0.64
–
Experimental measurements
Breath duration (s)
4.03±0.97
2.41–6.44
TABLE I : Breathing characteristics and sensing design targets. Measurements show mean ± SD and observed range.
Fig. 2 : Representative dorsoventral chest displacement trajectories recorded with OptiTrack during (a) normal breathing, (b) coughing, (c) hiccups, and (d) talking.
Fig. 3 : Assembly (a) shows the testbed used to emulate CT bore robot motion and to run tests within a controlled system. The design features a linear rail with two carriages driven by motorized capstan drives. Attached to the breathing model drive is a replaceable skin phantom that can be punctured. The needle insertion mechanism is used to control and drive the needle motion and has a clamp-based mechanism to allow the needle to be fixed and also exchanged after use. Assemblies (b) and (c) show a zoomed view of the contact sensors for the testbed and CT bore robot respectively, each of which read through a Hall effect sensor. Assembly (d) shows the assembly on the CT bore robot. The open circles in (b) and (c) indicate the point of contact on the contact sensor.
Fig. 4 : The gated needle insertion process involves four steps: (1) The robot base and wrist bring the needle to a position suitable for insertion. (2) The contact sensor is brought within range of the torso to maintain contact and sense breathing. (3) Once the breathing motion is captured, the needle is driven into the skin. (4) The needle is allowed to passively move with the breathing motion, applying a gated insertion when the suitable breathing stage is detected.
Name
Peak-to-Peak (mm)
SD (mm)
ToF Sensor
8
2
Contact Sensor
0.011
0.002
TABLE II : Sensor Noise
Fig. 5 : Comparison of the EKF forecast (red) and raw sensor values (gray) overlaid with the ground truth breathing motion (green) shows that the EKF forecast successfully adjusts for lag in sensor readings and closely tracks the ground truth.
Fig. 6 : The torso phantom and the measured breathing signal are shown in green and gray above respectively. In (a) the background colors indicate the phase of the sequence. Gray indicates initialization. Red indicates the needle carriage approaching. Orange indicates contact incremental pushes to reach full breath detection. Green indicates full breath detection and EKF settling. Blue slices indicate motions of gated insertion. The gated insertion segment is further detailed in (b), showing insertion highlighted by blue during full exhales.
Computed tomography (CT)-guided needle biopsies are critical for diagnosing a range of conditions, including lung cancer, but present challenges such as limited in-bore space, prolonged procedure times, and radiation exposure. Robotic assistance offers a promising solution by improving needle trajectory accuracy, reducing radiation exposure, and enabling real-time adjustments. In our previous work, we introduced a robotic platform designed for accurate needle insertion within the confined CT bore. However, its performance in clinical settings is restricted by limited dexterity and a constrained workspace. In this study, we present an 11-degree-of-freedom (DOF) robotic system that integrates a 6-DOF robotic base with an improved 5-DOF cable-driven end-effector, yielding a significantly expanded workspace and enhanced dexterity. To leverage the hyper-redundant degrees of freedom, we introduce a weighted inverse kinematics controller, along with a null-space control strategy to optimize maneuverability and dexterity. By using a task-oriented weight matrix as a hyperparameter, the system provides a two-stage priority scheme fit for both large-scale movement and fine in-bore adjustments. In clinically relevant simulated scenarios, the system demonstrates a consistent 97% reachability rate across various human models. In addition, the task-oriented weight-matrix policy is extensively explored in five representative subtasks seen during needle biopsy through both simulation and real-world experiments, demonstrating superior tracking accuracy and enhanced manipulability for CT-guided procedures.
Peihan Zhang, Derek Chen, Ishan Duriseti +7
Jacobs School of Engineering, University of California San Diego, La Jolla, CA 92093 USA · School of Medicine, University of California San Diego, La Jolla, CA 92093 USA · School of Medicine, University of Missouri-Kansas City, Kansas City, MO 64110 USA
Bronchoscopic navigation relies on registering endoscopic video to a preoperative CT scan, but respiratory motion deforms the airway by 5-20 mm, creating CT-to-body divergence that limits localization accuracy. In practice, this is mitigated through breath-hold protocols, which attempt to match the intraoperative anatomy to a static CT, but are difficult to reproduce and disrupt clinical workflow. We propose to eliminate the need for breath-hold protocols by leveraging patient-specific respiratory modeling. Paired inhale-exhale CT scans, already acquired for planning, implicitly define the patient-specific deformation space of the breathing airway. By registering these scans, we reduce respiratory motion to a single scalar breathing phase per frame, constraining all reconstructions to anatomically observed configurations. We embed this representation within a mesh-anchored Gaussian splatting framework, where a lightweight estimator infers breathing phase directly from endoscopic RGB, enabling continuous, deformation-aware reconstruction throughout the respiratory cycle without breath-holds or external sensing. To enable quantitative evaluation, we introduce RESPIRE, a physically grounded bronchoscopy simulation pipeline with per-frame ground truth for geometry, pose, breathing phase, and deformation. Experiments on RESPIRE show that our approach achieves geometrically faithful reconstruction, over 20x faster training, and 1.22 mm target localization accuracy (within the 3mm clinically relevant tolerances) outperforming unconstrained single-CT baselines. Please check out our website for additional visuals: https://asdunnbe.github.io/RESPIRE/
Andrea Dunn Beltran, Daniel Rho, Aarav Mehta +5
University of North Carolina at Chapel Hill · 2Harvard Medical School · University of California, San Diego
Percutaneous dilatational tracheostomy (PDT) is frequently performed on patients in intensive care units for prolonged mechanical ventilation. The needle puncture, as the most critical step of PDT, could lead to adverse consequences such as major bleeding and posterior tracheal wall perforation if performed inaccurately. Current practices of PDT puncture are all performed manually with no navigation assistance, which leads to large position and angular errors (5 mm and 30 degree). To improve the accuracy and reduce the difficulty of the PDT procedure, we propose a system that automates the needle insertion using a velocity-controlled robotic manipulator. Guided using pose data from two electromagnetic sensors, one at the needle tip and the other inside the trachea, the robotic system uses an adaptive constrained controller to adapt the uncertain kinematic parameters online and avoid collisions with the patient's body and tissues near the target. Simulations were performed to validate the controller's implementation, and then four hundred PDT punctures were performed on a mannequin to evaluate the position and angular accuracy. The absolute median puncture position error was 1.7 mm (IQR: 1.9 mm) and midline deviation was 4.13 degree (IQR: 4.55 degree), measured by the sensor inside the trachea. The small deviations from the nominal puncture in a simulated experimental setup and formal guarantees of collision-free insertions suggest the feasibility of the robotic PDT puncture.
Yuan Tang, Bruno V. Adorno, Brendan A. McGrath +1
Manchester Centre for Robotics and AI and the Department of Mechanical and Aerospace Engineering, University of Manchester, Manchester, UK · Manchester Centre for Robotics and AI and the Department of Electrical and Electronic Engineering, University of Manchester, Manchester, UK · Manchester University Foundation Trust Wythenshawe Hospital Intensive Care Unit, Manchester, UK