Supervised Autonomous Bimanual Suction and Cauterization for Bleeding Control in Low Grade Splenic Trauma
Authors: Francis Fan, Mackenzie A Eagleson, Yvonne Zhang, Hongting Pan, Lujia Yang, Ethan Kilmer, Jiawei Ge, Nural Yilmaz, Avinash Dhimal, Ahmed Ghazi, Jeffrey Jopling, Laura Connolly, Axel Krieger
Organizations: Laboratory of Computational Sensing and Robotics (LCSR), Johns Hopkins University, Baltimore, MD 21218, USA · Department of Surgery, Johns Hopkins Medicine, Baltimore, MD 21287, USA · Brady Urological Institute, Johns Hopkins University, Baltimore, MD 21287, USA
Uncontrolled hemorrhage remains a leading cause of preventable trauma mortality. Timely intervention is not always available, making autonomous systems that can aid in bleeding control a promising alternative. Given the safety-critical nature of the task, supervised autonomy, in which the system executes the technical steps while a human retains oversight, represents a necessary intermediate step toward this goal. In this work, we present a supervised autonomous bimanual robotic framework for injury cauterization (eletrosurgical coagulation) in low-grade splenic trauma. The system integrates real-time bleeding segmentation, markerless tool keypoint tracking, and deep tool-tissue contact detection to coordinate two robotic arms for synchronized suction and cauterization. Evaluated on a novel, cannulated ex vivo splenic hemorrhage model, the system achieved an average bleeding surface cauterization coverage of 82.9% across seven distinct injuries. This work demonstrates the feasibility of using a supervised autonomous bimanual system to deliver targeted coagulation across an injury site.
Figures & tables
Fig. 1: Supervised autonomous bimanual suction and cauterization for bleeding control. Images progress in time from top to bottom. Top: Blood pooling from a simulated splenic injury. Middle: Autonomous suction clearing blood from the injury site. Bottom: Autonomous cauterization covering the injury surface to initiate bleeding control.
Fig. 2: System diagram: Left: Cannulated vessels in the ex vivo splenic hemorrhage model. Middle: Experimental setup including suction pump, robotic tools, and perfusion system. The splenic hemorrhage model is outlined in green; robotic instrument positions are highlighted in yellow and pink. Right, top: Close-up of the yellow-outlined tool, a custom-designed robotic suction tool fitted to ProGrasp forceps. Right, bottom: Close-up of the pink-outlined tool, a cautery hook instrument used for wound coagulation.
Fig. 3: Overview of the supervised autonomous bimanual suction and cauterization workflow. The bleeding segmentation model is run on a laparoscopic image to produce a bleeding segmentation mask, from which the injury location is derived. The positions of both robotic tools (suction and cautery) are extracted from the image using a markerless tool-tracking framework based on keypoint detection. This tracking information is used to crop the image to a smaller region of interest for input to the tool–tissue contact detection model. The robotic tools are moved to the injury location via visual servoing, lowered until contact is detected, and then used to execute an autonomous Cartesian trajectory along the wound length.
Fig. 4: Images of injuries before and after suction and cauterization. Top row: Injury before any intervention. Middle row: Injury after cauterization. Bottom row: Overlap between the injury mask and the cauterized surface area.
Injury
Grade
Depth
Length
Coverage
(mm)
(mm)
(%)
1
I
10
7.4
98.69
2
I
10
7.8
98.74
3
I
10
8.4
67.89
4
I
10
5.2
67.85
5
I
10
10.0
100
TABLE I: Results of supervised autonomous bimanual suction and cautery. Column 1 is the injury ID, corresponding to Figure 4 . Columns 2 - 4 represent the extend of the injury (grade, length). Column 5 is the percentage of bleeding surface area cauterized.
Laboratory for Computational Sensing and Robotics, Johns Hopkins University, Baltimore, Maryland · Department of Pathology, Johns Hopkins Hospital, Baltimore, Maryland
Department of Electrical Engineering and Computer Sciences, University of California, Berkeley, CA 94720, USA. · Department of Robotics, Perception and Learning, KTH Royal Institute of Technology, Stockholm, SE 10044, Sweden. · Intuitive Surgical, Sunnyvale, CA 94086, USA.