ACUS: An Angular Calibrated Uncertainty Score for Incomplete-View CT

Authors: Zihan Deng, Chuanzhi Xu, Quankeng Huang, Yujie Liu, Zhenzhen Gong, Junjie Huang, Tianye Niu, Lequan Yu

Organizations: The University of Hong Kong · The University of Sydney · Chongqing University · University of Science and Technology of China · 

Abstract

Sparse-view and limited-arc computed tomography (CT) are commonly treated as projection completion: a network fills an unmeasured view, and the result is ranked by similarity to one ground-truth image. That score can be high even when the query angle is far from every acquired measurement, because a learned prior may invent anatomy the scan never constrained. Existing protocols therefore leave a basic question unanswered: how does predictive reliability change with distance from the acquired views? We propose the Angular Calibrated Uncertainty Score (ACUS) to answer that question. ACUS is the Winkler interval score of a nominal 90% predictive interval on held-out projections, binned by distance to the nearest acquired view. It charges interval width and adds a large miss penalty when a held-out pixel falls outside the interval, so a narrow band helps only when coverage holds. We use collapse on acquired angles as a necessary consistency check on the predictor. On chest and dental cone-beam orbits, interval width grows with the angular gap. Empirical coverage ranges from 77% to 92%, and the peak signal-to-noise ratio (PSNR) of the point estimate falls as the width rises. Nearest-neighbor, interpolation, and point completion, each given one residual scale, have zero reported far coverage and incur outside-distance penalties. ACUS exposes gap-dependent overconfidence that a point-estimate PSNR alone does not measure.

Figures & tables

Appendix figures & tables33 assets

Supplementary material from the paper’s appendix.

Appendix

Explore similar work

Jul 15, 2026

Posterior Variance Is a Constraint Map, Not an Error Map: Closed-Form Uncertainty for Radiative Gaussian Splatting in Sparse-View CT

Radiative Gaussian splatting reconstructs sparse-view CT fast and accurately, and recent work attaches per-Gaussian posteriors to yield per-voxel uncertainty maps. We ask what such a map actually measures: posterior variance is a data-constraint map, not an error map -- its alarms are trustworthy, its all-clears are not. Exploiting the strict linearity of X-ray rendering in the per-Gaussian densities, we derive a clamp-aware closed form that the unchanged rasterizer evaluates exactly in one forward pass, in volume and projection space: the infinite-sample limit of the sampling estimator of concurrent work, at ~8x lower cost. On the official 15-scene benchmark this uncertainty ranks true error on 14 of 15 scenes. Restricted to the object interior -- the tissue a clinician reads -- the ranking collapses (median Spearman 0.11, 0/15 pass), identically for a deep ensemble and for a strictly positive log-normal posterior: three constructions, two estimator families, no survivors. The mechanism is structural: about 90% of in-object error is bias that reproduces across retrainings, invisible to model disagreement; 73-81% of the full-volume correlation is carried by object/surround contrast; and an exactly solvable control puts the observed in-object ranking 4-5x below what a perfectly calibrated posterior with the same sigma-spread would score. The error scale, by contrast, is an engineering problem, and we solve it: reparameterizing the posterior contracts the cross-scene temperature spread from 19.3x to 2.6x, one scene-agnostic temperature transfers to unseen scenes (10/15 leave-one-scene-out), and the repaired scale tracks photon count at the Poisson-predicted -1/2 power. We distill evaluation practice that would have caught the illusion -- masked calibration, seed-wise bias decomposition, an exact-posterior reference -- and release all protocols, seeds and per-run evidence.
Oct 5, 2026

Geometry-Aware Diffusion Approximate Posterior Sampling for Sparse-View and Limited-Angle CT

Sparse-view computed tomography (CT) reduces radiation dose and acquisition time and may mitigate motion artifacts. However, angular undersampling provides insufficient information to determine the image uniquely and stably. Limited-angle CT, arising from restricted angular coverage, produces strongly directional information loss associated with the missing angular range. In both settings, image directions may be strongly observed, weakly constrained, or unobservable, leading to severe ill-posedness and reconstruction ambiguity. Existing diffusion-based approaches incorporate measurement information through likelihood guidance, data-consistency operations, or range-null-space corrections. However, they do not generally use the continuously varying measurement sensitivity of the acquisition to jointly shape both reconstruction updates and stochastic exploration. We propose a geometry-aware diffusion-guided stochastic reconstruction framework for sparse-view and limited-angle CT. Its central component is a regularized noise-weighted pullback metric constructed from the CT forward operator and measurement-noise covariance. This metric continuously adapts both the measurement-aware update and stochastic exploration according to directional measurement sensitivity, suppressing changes along strongly constrained directions while permitting greater exploration along weakly constrained and unobservable directions. We complement this geometry-aware update with a regularized data-consistency correction and approximately null-space-restricted stochastic perturbations, implemented matrix-free using forward and backprojection operations together with conjugate-gradient solves. Experiments on sparse-view, noisy, and limited-angle CT demonstrate competitive reconstruction quality, strong measurement consistency, and spatially resolved empirical uncertainty estimates.
Sep 12, 2026

SONAR: A Structure-Consistent Neural Operator for Null-Space-Aware Sparse View CT Reconstruction

Sparse-view computed tomography (CT) reduces radiation dose and acquisition time but remains severely ill-posed because incomplete projections poorly constrain null-space information. Existing learning-based methods often estimate this information in high-dimensional image space, conflate physical measurement errors with prediction errors, and depend on fixed discretizations. We propose SONAR, a Structure-Consistent Neural Operator for Null-Space-Aware Reconstruction. Instead of recovering the full null-space component, SONAR predicts a low-dimensional null-space-aware representation from the acquired projections as pseudo-measurements. It separates measurement and pseudo-measurement residuals, lifts them into the image domain through physics operators, and applies independent neural operators to constrain their structural effects, thereby accommodating admissible errors while suppressing unsupported structures. To support cross-discretization reconstruction, an anisotropic U-shaped neural operator models the periodic angular and nonperiodic detector dimensions using direction-dependent continuous supports, while image-domain neural operators re-discretize continuous kernels on target grids. These components form an optimization-inspired unrolled network. Experiments on simulated AAPM and clinical MARS photon-counting CT data demonstrate consistent improvements across seen and unseen view settings and unseen image resolutions. On AAPM dataset, SONAR improves PSNR by 1.87dB at 62 views and by 7.63dB under zero-shot transfer to a 512×512512\times512 grid over the strongest competing methods. SONAR also achieves the best overall performance in all clinical settings evaluated, demonstrating accurate, structurally reliable, and discretization-robust sparse-view CT reconstruction.