Personalization of a Commercial Powered Knee Prosthesis Under Clinically Realistic Conditions via Reinforcement Learning: A Feasibility Study
Authors: Xinyi Liu, Atli Örn Sverrisson, David Langlois, Ming Liu, Varun Nalam, He Huang
Organizations: Lampe Joint Department of Biomedical Engineering, University of North Carolina at Chapel Hill and North Carolina State University, Raleigh, NC, USA.
Personalization of powered prosthesis control to individual user needs remains a significant challenge. To address this challenge, policy iteration with constraint embedded (PICE), a reinforcement learning (RL) tuning approach, has been developed and validated. This feasibility study investigated whether the developed approach can personalize a commercial powered knee prosthesis under clinically realistic conditions. To reduce the time for personalization, tuning was initialized with a policy learned offline from one non-disabled participant walking on a treadmill and was then applied to three other non-disabled participants walking with the knee through a bypass adapter on a 5-meter walkway. All participants reached the goal within 8-18 tuning steps (1.7-4.2 minutes of walking), reducing the knee-kinematic feature error by 64-90% on average. These results support the feasibility of implementing the RL-based tuning approach in clinically realistic conditions with space and equipment constraints.
Figures & tables
Fig. 1: Conceptual overview of prior-informed prosthesis personalization in an envisioned clinical workflow. A clinician specifies a target set of gait features, and sensor measurements from the prosthetic knee are used to quantify deviations from the prescribed target. PICE initializes a tuning policy π using prior experience Do and updates the policy using user-specific tuning experience Dnew collected during walking. The updated policy generates parameter adjustments Δc for the proprietary low-level controller, enabling iterative personalization through a closed-loop tuning process without replacing the underlying controller. Red and blue indicate onboard and offboard computation, respectively.
Fig. 2: Experimental setup and parameterization. Left: participant wearing the Össur Power Knee, with a wire streaming data from the knee. Top right: the four clinical parameters tuned by the policy, c1 initial stance angle, c2 stance knee flexion level, c3 swing initialization level, and c4 maximum swing flexion angle. Bottom right: knee angle over two gait cycles, streamed from the knee’s onboard sensors at about 50 Hz; shaded regions mark stance.
Age
Height
Mass
Target y∗
ID
Role
Sex
(yr)
(m)
(kg)
T1
T2
AB0
Prior
M
30
1.73
75.0
AB1
Tuning
M
21
1.70
68.0
✓
AB2
Tuning
M
30
1.76
86.2
✓
✓
AB3
Tuning
M
33
1.70
70.3
✓
TABLE I: Study participant information and experimental design.
Fig. 3: Online tuning in the four trials: AB1, AB2, and AB3 with target T1 (solid lines) and AB2 with target T2 (dashed line). (A) Reward at each tuning (RL) step. (B) Root-mean-square (RMS) value of temporal-difference (TD) error ( Eq. 4 ) of AB3’s learned action-value function on the held-out steps 13–18, for the offline policy and the policy after the first online update (online). (C) Error of each knee-kinematic feature from its target, yi−yi∗ , in degrees. Gray bands mark the goal range: a reward above −0.5 in (A) and the goal threshold of each feature in (C). Vertical dashed lines in (A) and (C) mark step 12, the first scheduled online policy update ( Nb=12 ), which only AB3 reached.
Fig. 4: Start versus end of tuning for the four trials. (A) Reward at the first (light) and last (dark) tuning step; arrows show the change and percentages the improvement. (B) Net walking time until the success criterion was met. (C) Absolute error of each knee-kinematic feature, ∣yi−yi∗∣ , at the first (light) and last (dark) tuning step; dashed lines mark the goal threshold of each feature, and percentages give the reduction in the error averaged over the four trials. Hatched bars: AB2 with target T2.
Fig. 5: Knee angle over the gait cycle at the start of tuning (first four gait cycles, dashed black) and at the end (last tuning step, colored), with the target of each feature (gray bars); shading shows the spread across gait cycles.
Reinforcement learning control uses reward-driven learning to make decisions in dynamic systems, from robots and vehicles to energy and infrastructure.
Woodruff School of Mechanical Engineering and the Institute for Robotics and Intelligent Machines, Georgia Institute of Technology, Atlanta, GA 30332-0405 USA · Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA 30332-0405 USA
Division of Industrial Data Science (DIDS), School of Data Science, Lingnan University, Hong Kong · Department of Mechanical and Energy Engineering, Southern University of Science and Technology, Shenzhen, 518055, China