Organizations: Chair for Computer Aided Medical Procedures(CAMP), Technical University of Munich(TUM), Munich, Germany · Munich Center for Machine Learning (MCML), Munich, Germany
Accurate needle placement in spine interventions is critical for effective pain management, yet it depends on reliable identification of anatomical landmarks and careful trajectory planning. Conventional imaging guidance often relies both on CT and X-ray fluoroscopy, exposing patients and staff to high dose of radiation while providing limited real-time 3D feedback. We present an optical see-through augmented reality (OST-AR)-guided robotic system for spine procedures that provides in situ visualization of spinal structures to support needle trajectory planning. We integrate a cone-beam CT (CBCT)-derived 3D spine model which is co-registered with live ultrasound, enabling users to combine global anatomical context with local, real-time imaging. We evaluated the system in a phantom user study involving two representative spine procedures: facet joint injection and lumbar puncture. Sixteen participants performed insertions under two visualization conditions: conventional screen vs. AR. Results show that AR significantly reduces execution time and across-task placement error, while also improving usability, trust, and spatial understanding and lowering cognitive workload. These findings demonstrate the feasibility of AR-guided robotic ultrasound for spine interventions, highlighting its potential to enhance accuracy, efficiency, and user experience in image-guided procedures.
Figures & tables
Fig. 1 : AR-Guided Needle Insertion. The system integrates robotic ultrasound, CBCT, and AR visualization into a unified framework, providing intuitive guidance during needle insertion for spine procedures.
Fig. 2 : System Overview. The pipeline integrates robotic CBCT, robotic ultrasound, and AR into a unified framework, enabling in situ AR visualization of spine anatomy and needle guidance. ( ∙ ) represents CBCT volume acquisition, ( ∙ ) depicts ultrasound and robot state acquisition, and ( ∙ ) indicates the visualization process.
Fig. 3 : Transformation Chain for System Registration. Note that the transformations shown with solid arrows are acquired, while the transformations with dashed arrows are derived. ( ∙ ) represents transformations in the virtual domain, ( ∙ ) corresponds to transformations in the real world, and ( ∙ ) denotes the registration between virtual and real robot.
Fig. 4 : AR Guidance Visualization. Raycasting from the needle guide onto the CT-derived spine model yields the predicted intersection point (marked in red) between the needle and the spine model.
Fig. 5 : Study Setup. Experimental configuration with robotic arm, ultrasound probe, lumbar spine phantom, and screen-based guidance interface.
Achieving optimal screw placement for orthopedic surgeries requires frequent alignment checks and multiple anatomical views under X-ray -- a process known as "fluoro-hunting" that increases radiation exposure to patients and surgical teams. This work introduces X-GuideAR, an augmented reality (AR) framework for identifying optimal X-ray views, aimed at reducing radiation exposure while ensuring accurate screw placement. To exemplify the benefits of X-GuideAR, we focus on S2 alar-iliac (S2AI) screw placement. Our system provides radiation-free guidance for view acquisition and drilling by generating synthetic X-ray previews that accelerate fluoro-hunting. Once the required anatomical views are identified using these previews, a real X-ray is acquired, and the preview of the drilling trajectory is augmented onto it, facilitating precise screw placement with minimal additional radiation. A preliminary study involving eight S2AI trajectories performed by an expert spine surgeon demonstrated a 62.3% reduction in the number of X-rays. Post-procedure evaluations showed that trajectories done with X-GuideAR supported an average safe screw diameter of 12.95 mm compared to 5.9 mm under the conventional workflow, suggesting improved bony containment and potential biomechanical benefit. X-GuideAR shows great potential to reduce radiation exposure and streamline S2AI screw placement, offering a promising direction toward safer and more efficient surgeries.
Mingxu Liu, Zixuan Liu, Ruchen Cai +5
Department of Computer Science Johns Hopkins University Baltimore, MD, 21211, USA · Department of Orthopaedic Surgery Johns Hopkins University Baltimore, MD, USA · Department of Orthopaedic Surgery Johns Hopkins School of Medicine Baltimore, MD, USA +3
Intraoperative Cone Beam Computed Tomography (CBCT) provides a reliable 3D anatomical context essential for interventional planning. However, its static nature fails to provide continuous monitoring of soft-tissue deformations induced by respiration, probe pressure, and surgical manipulation, leading to navigation discrepancies. We propose a deformation-aware CBCT updating framework that leverages robotic ultrasound as a dynamic proxy to infer tissue motion and update static CBCT slices in real time. Starting from calibration-initialized alignment with linear correlation of linear combination (LC2)-based rigid refinement, our method establishes accurate multimodal correspondence. To capture intraoperative dynamics, we introduce the ultrasound correlation UNet (USCorUNet), a lightweight network trained with optical flow-guided supervision to learn deformation-aware correlation representations, enabling accurate, real-time dense deformation field estimation from ultrasound streams. The inferred deformation is spatially regularized and transferred to the CBCT reference to produce deformation-consistent visualizations without repeated radiation exposure. We validate the proposed approach through deformation estimation and ultrasound-guided CBCT updating experiments. Results demonstrate real-time end-to-end CBCT slice updating and physically plausible deformation estimation, enabling dynamic refinement of static CBCT guidance during robotic ultrasound-assisted interventions. The source code is publicly available at https://github.com/anonymous-codebase/us-cbct-demo.
Feng Li, Ziyuan Li, Zhongliang Jiang +2
Chair for Computer-Aided Medical Procedures and Augmented Reality, Technical University of Munich, Munich, Germany · Munich Center for Machine Learning, Munich, Germany · The University of Hong Kong, Hong Kong, China
The handling and assembly of instruments during surgery imposes high cognitive demands on scrub nurses, particularly when instruments are unfamiliar. We present a supporting guidance system for surgical instrumentation that combines multi-camera 6D pose estimation with augmented reality in-situ visualization on a head-mounted display without the requirement for additional markers. Pose estimation and consecutive camera calibration are achieved through known objects. The 6D pose estimation network is trained purely on synthetic data, aiming for better generalizability and real-world applicability. The AR guidance displays tooltip localization cues and step-wise assembly animations. Via gaze-based selection and a foot pedal, users can switch between assembly steps in intraoperative use. In a technical evaluation, our approach outperforms state-of-art 6D pose estimation. A user study with 29 scrub nurses was conducted in a surgical simulation of knee arthroplasty, comparing the system against a paper manual. AR guidance significantly reduced the perceived workload compared. Objectively, AR guidance reduced task completion time by 21.3% (4.76 minutes). Specifically, scrub nurses less experienced with the instrument set benefited when using the system. Error frequencies were comparable between conditions. Qualitative feedback highlighted improved process clarity, reduced information overload, and perceived independence. To summarize, our marker-free multi-camera AR guidance approach for surgical instruments can, subjectively and objectively, improve intraoperative instrumentation performance, particularly for untrained scrub nurses.
Shiyu Li, Julian Kreimeier, Hannah Schieber +4
Technical University of Munich, Human-Centered-Computing and Extended Reality Lab · TUM University Hospital, Orthopedics and Sports Orthopedics · Munich Institute of Robotics and Machine Intelligence (MIRMI) +2