A Transparent 3-DoF Assistive Device for Upper-Limb Tremor and Ataxia Compensation: Design and Experimental Validation
Authors: Thomas Dobosz, Veronika Hofmann, Ahmed Barhoma, Ann-Kathrin Henke, Johannes A. Gaus, Hans Mees, Winfried Ilg, Isabell Wochner, Martin Giese, Andreas Bulling, Syn Schmitt, Christophe Maufroy, Urs Schneider, Thomas Bauernhansl, Daniel Haeufle
Organizations: Institute of Industrial Manufacturing and Management, University of Stuttgart, Germany · Biomechatronic Systems, Fraunhofer IPA, Stuttgart, Germany · Hertie Institute for Clinical Brain Research and Centre for Integrative Neuroscience, University of Tubingen, Germany · Center for Bionic Intelligence Tubingen Stuttgart (BITS), Germany · Institute for Visualization and Interactive Systems, University of Stuttgart, Germany · Institute for Modelling and Simulation of Biomechanical Systems, University of Stuttgart, Germany
Assistive devices for patients with neurodegenerative movement disorders must combine high mechanical transparency with the ability to generate corrective interaction forces. Achieving both properties simultaneously remains challenging. This paper presents the completed iAssistADL platform, a 3-DoF assistive device for future tremor and ataxia compensation during activities of daily living. The system employs a transparency-oriented mechanical architecture with all actuators located at the base, resulting in low moving mass, decoupled kinematics, and high mechanical transparency. To compensate for friction-induced transparency limitations along the vertical axis, a hybrid control strategy is proposed that combines open-loop force rendering in the horizontal plane with closed-loop force control in the vertical direction. The platform is experimentally evaluated through transparency measurements and static and dynamic force-rendering experiments at representative configurations. Results demonstrate resistive torques below 0.1 Nm in the horizontal plane, end-effector forces of up to approximately 10 N throughout the workspace, and accurate horizontal force rendering up to 4 Hz. While the vertical axis remains limited by friction-induced dynamics at higher frequencies, the proposed system successfully combines high transparency with controlled force generation, establishing a promising hardware platform for future tremor and ataxia compensation.
Figures & tables
Fig. 1 : User interacting with the iAssistADL platform. The device applies assistive forces to the forearm while preserving natural upper-limb motion.
Fig. 2 : iAssist prototype along with the power supply and real-time computer. Prototype end-effector connected to a 6-D force sensor on a rigid fixture.
Fig. 3 : Schematic of the assistive device with three active DoF. A prismatic joint along the z-axis (blue) actuated via a scissor mechanism, whose motion is constrained by the ball-spline mechanism, a first rotational joint along the ball-spline axis (red), and a second rotational joint bilaterally actuated using cables routed through the scissor mechanism (green). A passive spring mechanism (bottom left) is used to balance the weight of the arm. With joint angles θ1,θ2 , θ3 and motor torques τ1 , τ2 . τ3 . Adopted from [ 18 ]
Force capability
Point
(x,y,z) [cm]
θ1
θ2
θ3
fz,max [N]
Guar. ∥fxy∥ [N]
Max ∥fxy∥ [N]
Anisotropy
detJxy
P1 grasp bottle
(0,50,30)
33.75∘
146.25∘
30.00∘
10.5
9.89
18.2
1.84
0.187
P2 fill glass
(−10,60,25)
52.00∘
146.93∘
24.62∘
10.0
9.89
15.0
1.51
0.202
P3 to mouth
(−10,30,30)
39.00∘
177.86∘
30.00∘
10.5
9.89
28.8
2.91
0.133
TABLE I : Operating points, inverse-kinematic solution, and force-capability analysis. The guaranteed omnidirectional planar force is the shoulder-torque–limited lower bound min(a1,a2) and is configuration-independent : it holds over the entire reachable workspace, not only at P1 – P3 . The pose-dependent quantities are the vertical capability fz,max , the maximum planar force (parallelogram diagonal), the anisotropy (max/guaranteed), and the Jacobian manipulability detJxy .
Fig. 4 : Test positions with glass (G) and bottle (B). F = Front, B = Bottom, L = Left, R = Right.
Metric
Fx
Fy
Fz
Rise time tr (10–90 %)
16 ms
15 ms
3.04 s
Overshoot
33%
33%
28%
Settling time
1.35 s
1.24 s
6.95 s
Steady-state error
−0.04 N
0.08 N
−0.12 N
(relative)
2.1%
5.5%
−16%
TABLE II : Step-response metrics (mean over all cases).
f
Metric
Fx
Fy
Fz
2 Hz
Ar
0.91
0.92
0.98
Δt (ms)
142
155
191
RMSE ϕ (N)
0.36
0.40
1.24
4 Hz
Ar
0.95
0.97
1.87
Δt (ms)
127
124
176
RMSE ϕ (N)
0.68
0.57
3.37
TABLE III : Sinusoidal tracking (mean): amplitude ratio Ar , transport delay Δt , phase-compensated RMSE.
Fig. 5 : Open-loop step response of the Fx axis. Reference (dashed) versus measured force (solid); shaded region is the ±5% settling band.
Fig. 6 : Stick–slip limit cycle on the vertical axis for the Fz=5 N step reference at pose P3. The commanded control torque (top) ramps under integral action while the ball-spline carriage is stuck, breaks away and overshoots on slip, then re-sticks at a new offset, producing a sustained oscillation of the measured force (bottom) about the reference (dashed). The series compliance of the lightweight distal links stores and discharges elastic energy on each slip event, so the integrator hunts across the stiction dead-band without converging to a unique equilibrium.
Fig. 7 : Sinusoidal force tracking on the horizontal Fx axis. Top: the commanded reference (dashed), the raw measured force (grey), and the phase-compensated response (blue). The raw signal reproduces the reference amplitude but is shifted by the near-constant transport delay ( ≈140 ms), which at this frequency manifests as a large phase lag; removing this delay recovers the phase-compensated trace, which closely overlays the reference. Bottom: instantaneous tracking error before (grey) and after (blue) phase compensation, showing that the raw error is dominated by the pure time delay—its removal collapses the error to the residual amplitude- and shape-tracking component (phase-compensated RMSE =0.36 N at 2 Hz), showing that the raw RMSE is dominated by the measured transport delay.
School of Information and Software Engineering, University of Electronic Science and Technology of China, Chengdu 610054, Sichuan, P.R. China · Sichuan Provincial Key Laboratory for Human Disease Gene Study, Sichuan Academy of Medical Sciences & Sichuan Provincial People’s Hospital, University of Electronic Science and Technology of China, Chengdu, Sichuan, China. · Research Unit for Blindness Prevention, Chinese Academy of Medical Sciences, Sichuan Academy of Medical Sciences and Sichuan Provincial People’s Hospital, Chengdu, Sichuan, China.