Controlling scientific instruments requires parameters to be adjusted progressively in response to feedback from successive operations. When unobserved operating conditions evolve with operational history, responses become nonstationary, and a static mapping from actions to responses becomes inadequate for decisions across shots. Shot-to-shot tuning of neutral beam injection (NBI) ion sources in magnetic confinement fusion devices is representative of this class of problems: the conditions of electrode surfaces and vacuum are difficult to observe directly, discharge outcomes vary between shots, and experiments are costly and carry operational risks. We propose a physically grounded multimodal agent with three independently auditable components. Deterministic algorithms extract symbolic physical quantities and visual evidence from multichannel waveforms; a memory with two layers organizes trajectories from the current task, experience from earlier tasks, and procedural knowledge, from which a multimodal large language model (MLLM) generates candidate actions; and an independent action feasibility check enforces action constraints before execution. In a strictly chronological replay of 455 shots from four EXL-50U platforms, physical symbols alone outperformed waveform images alone by 0.21-0.32 in macro-F1 for quality classification across all four evaluated models. Retrieved context increased the mean agreement with historical action directions by 22.4-42.7 percentage points, whereas the contribution of visual evidence varied across models. Under supervised operation on five ion sources, the system achieved a beam-extraction rate of 86.80% across 409 executed shots. These results demonstrate the engineering feasibility of the framework for supervised sequential tuning on a physical scientific instrument.
Figures & tables
Figure 1: Measured multichannel waveforms from three consecutive shots (shots 22541–22543) under identical control settings. Although all five control variables were held fixed, the recorded acceleration currents were 41.54 A, 0.00 A, and 33.65 A, respectively. The lower states z1 – z3 schematically represent unobserved operating conditions and denote history-dependent factors, beyond the current control variables, that may affect the response; the bottom annotation summarizes that the same action a yields different responses y as the latent state z drifts.
Local magnetic measurements and global shape objectives
Action bounds and device interlocks
Yes (plasma control)
Standard/safe Bayesian optimization [ 27 , 28 ]
Not addressed for latent-state drift
No
No
Point observations and a global surrogate
Conditional guarantees in safe variants
Applicable to online experiments
LLM agents for scientific experiments [ 31 , 32 ]
Through context or memory
No
Yes
Task-relevant inputs; no explicit integration
Can use semantic guardrails or external validators [ 38 ]
Yes (chemical experiments)
This work
History-conditioned dual-layer memory
No
Yes
Quantitative details and overall waveform morphology
Deterministic action feasibility check
Supervised shot-to-shot execution
Table 1: Qualitative comparison of the task conditions in this study with representative automated decision-making paradigms
Figure 2: Structure of the 50 kV hot-cathode arc-discharge positive ion source investigated in this study. The device consists primarily of a discharge-chamber backplate integrating the filaments and feedthrough assemblies, a multicusp-field discharge chamber in which the plasma is generated, and a three-electrode accelerator comprising the plasma, suppression, and ground grids.
Figure 3: Overall architecture of the physics-grounded multimodal agent. Deterministic signal grounding and annotated waveform rendering transform raw records into symbolic and visual evidence. Dual-layer memory supplies retrieved temporal context, which the MLLM integrates with the current evidence to propose a candidate action for the next shot. The candidate is passed unchanged to the PLC only after an independent feasibility check confirms every condition in Ct ; otherwise, it is rejected for regeneration or escalated to the operator. The executed shot is subsequently read back to update the archive and memory.
Figure 4: Symbolic grounding pipeline. After preprocessing and interval and event detection, the raw multichannel waveforms yield flat-top statistics, channel-missing flags, breakdown and oscillation events, and fault flags.
Figure 5: Typical acceleration-current waveform morphologies to be distinguished by the grounding operator: single flat top, multiple flat tops, beam oscillation, complete breakdown, and two types of sensor fault. The vertical axis of each panel is normalized separately and is intended only for morphological comparison.
Figure 6: Examples of annotations on measured acceleration-current waveforms: (a) beam flat-top identification for shot 11888; (b) high-frequency oscillation-dip identification for shot 12247. Dark gray denotes the raw waveform; the shaded band bounded by dotted lines marks the detected pulse interval, with the flat-top start ts , end te , mean current IˉACC , and width τ annotated in (a), and red triangles in (b) mark the detected oscillation points.
Class
Number of shots
Full breakdown
67
Insufficient pulse width
113
Long-pulse high current
80
Long-pulse low current
81
High-frequency oscillation
51
Fault noise
63
Table 2: Distribution of the six waveform classes in the 455-shot offline dataset
Configuration
Symbolic grounding
Visual evidence
Experience and knowledge retrieval
Feasibility check
Human feedback
Baseline
×
✓
×
✓
×
Physical-only
✓
×
×
✓
×
+Physical
✓
✓
×
✓
×
+RAG
✓
✓
✓
✓
×
Full
✓
✓
✓
✓
✓
Table 3: Five unified configurations for cross-model evaluation
Platform
Shots
Current MAE/A
Main-width MAE/ms
Flat-top-count accuracy
Flat-top event- count F 1
Breakdown-count accuracy
Breakdown event- count F 1
Test Bench 1
127
0.16
4.46
100.00%
100.00%
100.00%
100.00%
Test Bench 2
116
0.64
14.35
95.69%
96.08%
100.00%
100.00%
Experimental Bench 1
110
0.70
13.24
98.18%
97.44%
99.09%
97.56%
Experimental Bench 2
102
1.34
10.35
94.12%
96.20%
100.00%
100.00%
Pooled
455
0.68
10.42
97.14%
97.61%
99.78%
99.66%
Table 4: Independent validation of the symbolic grounding operator Φ on four platforms
Model
Configuration
Quality-classification macro-F 1
Current MAE/A
Pulse-width MAE/ms
Flat-top-count MAE
Mean directional agreement
Qwen3.5-Plus
Baseline
0.47
3.67
138.19
0.27
36.7%
Qwen3.5-Plus
Physical-only
0.79
1.40
28.37
0.09
54.4%
Qwen3.5-Plus
+Physical
0.73
1.54
42.23
0.09
55.8%
Qwen3.5-Plus
+RAG
0.71
1.24
19.36
0.09
78.2%
Qwen3.5-Plus
Full
0.68
1.31
24.33
0.08
86.0%
MiniMax3
Baseline
0.57
3.91
107.80
0.24
35.6%
Table 5: Overall results for four models under five configurations
Ion source
Executed shots
With beam extraction
Without beam extraction
Beam-extraction rate
Test Bench 2
49
45
4
91.84%
Source 1, Experimental Bench 2
64
53
11
82.81%
Source 2, Experimental Bench 2
83
69
14
83.13%
Source 3, Experimental Bench 2
103
91
12
88.35%
Source 4, Experimental Bench 2
110
97
13
88.18%
All
409
355
54
86.80%
Table 6: Summary of physical operation results for five ion sources
Figure 7: Comparison of shot-level beam-extraction rates between manual tuning and supervised agent tuning (AI tuning in the figure) for four ion sources. Numbers inside the bars denote shots with beam extraction and physically executed shots, and labels above the bars give the corresponding extraction rates. Error bars are 95% Wilson intervals calculated from shot-level binomial samples and characterize the sampling uncertainty of each beam-extraction rate. Source IDs 5–8 on the horizontal axis are the identifiers in the raw records and correspond, respectively, to Sources 1–4 of Experimental Bench 2 in Table 6 .
Figure 8: Shot-to-shot response trajectories from three consecutive experiments under the target-oriented protocol. Bars indicate beam pulse width, lines indicate the median acceleration current within the valid interval, and red crosses indicate shots in which no valid beam interval was detected.
Metric
Trial 1
Trial 2
Trial 3
Total
Shot range
11394–11402
11884–11891
11948–11953
—
Number of shots
9
8
6
23
Shots with a valid beam interval
8
7
6
21
Valid-beam-interval rate
88.89%
87.50%
100.00%
91.30%
Median acceleration-voltage range/kV
26.39–33.36
21.06–44.54
22.43–38.44
21.06–44.54
Median acceleration-current range/A
16.60–26.88
10.31–23.46
10.89–28.56
10.31–28.56
Table 7: Waveform statistics from three consecutive experiments on the same ion source at Test Bench 2. Voltage, current, and pulse width are reported as ranges of shot-level statistics among shots with a valid beam interval.
Appendix figures & tables6 assets
Supplementary material from the paper’s appendix.
Appendix
Figure 9: Offline performance of four models under five configurations. From left to right, the panels report field-derived six-class macro-F 1 , mean directional agreement across five action parameters, quality-classification macro-F 1 , and acceleration-current MAE.
Model
Configuration
Current/A
Pulse width/ms
Flat-top count
Quality classification
MAE
RMSE
MAE
RMSE
MAE
RMSE
Accuracy
F 1
Qwen3.5-Plus
Baseline
3.67
14.41
138.19
328.19
0.27
0.66
0.51
0.47
Qwen3.5-Plus
Physical-only
1.40
5.13
28.37
86.58
0.09
0.41
0.79
0.79
Qwen3.5-Plus
+Physical
1.54
7.08
42.23
121.89
0.09
0.34
0.73
0.73
Qwen3.5-Plus
+RAG
1.24
4.92
19.36
67.51
0.09
0.37
0.74
0.71
Qwen3.5-Plus
Full
1.31
5.05
24.33
77.34
0.08
0.36
0.84
0.68
Appendix
Table 8: Detailed recognition results for the four models. Continuous variables are reported as MAE/RMSE; quality classification is reported as Accuracy and macro-F 1 .
Model
Configuration
VACC
VFIL
VARC
Q
VDEC
Mean
Qwen3.5-Plus
Baseline
55.6%
47.9%
26.6%
18.7%
34.9%
36.7%
Qwen3.5-Plus
Physical-only
66.6%
69.2%
59.6%
39.8%
36.9%
54.4%
Qwen3.5-Plus
+Physical
66.6%
76.9%
56.9%
40.9%
37.8%
55.8%
Qwen3.5-Plus
+RAG
87.9%
82.2%
72.3%
75.4%
73.2%
78.2%
Qwen3.5-Plus
Full
89.7%
84.4%
85.1%
81.3%
89.7%
86.0%
MiniMax3
Baseline
49.2%
48.8%
23.7%
19.1%
37.4%
35.6%
Appendix
Table 9: Directional agreement (Accuracy) for five parameters across the four models. The mean is the arithmetic mean of the directional agreement for the five parameters.
Ion source
Manual tuning
Agent tuning
Difference (percentage points)
Beam extracted/ executed
Extraction rate
Beam extracted/ executed
Extraction rate
ID 5 (Source 1)
377/436
86.47%
53/64
82.81%
−3.7
ID 6 (Source 2)
337/419
80.43%
69/83
83.13%
+2.7
ID 7 (Source 3)
361/419
86.16%
91/103
88.35%
+2.2
ID 8 (Source 4)
346/421
82.19%
97/110
88.18%
+6.0
Total
1421/1695
83.83%
310/360
86.11%
+2.3
Appendix
Table 10: Comparison of shot-level beam-extraction rates between manual tuning and supervised agent tuning on four ion sources
Symbol
Physical quantity
Unit
PLC permissible range
Maximum per-shot change
VACC
Acceleration voltage
kV
5 – 45
Adjustment step: 1–5 kV
VFIL
Filament voltage
V
5 – 12
Adjustment step: 0.1–0.5 V
VARC
Arc voltage
V
70 – 130
Adjustment step: 1–5 V
Q
Gas flow
sccm
100 – 700
Adjustment step: 10–50 sccm
VDEC
Suppression voltage
V
0 – 3200
Constrained during the check to 4%–10% of VACC
Appendix
Table 11: Agent action parameters and PLC permissible constraints
Symbol
Meaning
Unit
Value and determination method
wmed
Median-filter window width
samples
15
3σ
Measurement-spike rejection bound
dimensionless
3 ; ∣x−medfilt∣ is compared with the global standard deviation, rather than a local baseline
α
Fraction trimmed from each end for the trimmed mean
%
30
wtrim
Sliding-window width for the trimmed mean
samples
21
θabs
Absolute-amplitude threshold for candidate pulses
A
1 ; the acceleration-current threshold is identical across all four platforms
rsel
Relative coefficient for candidate pulses
dimensionless
0.8 ; relative to the maximum pulse mean
Appendix
Table 12: Signal-processing thresholds of the grounding operator Φ