A Master-to-Simulation Interface with Demonstration-Trajectory Guidance for Vascular Intervention
Organizations: Beijing Institute of Technology, Beijing 100081, China · Department of Electronic and Electrical Engineering, Southern University of Science and Technology, Shenzhen, Guangdong 518055, China · School of Life Science and the Key Laboratory of Convergence Medical Engineering System and Healthcare Technology, Ministry of Industry and Information Technology, Beijing Institute of Technology, Beijing 100081, China · Shenzhen Institute of Advanced Biomedical Robot Company Limited, Shenzhen 518000, Guangdong Province, China ·
Abstract
Robotic vascular intervention can reduce radiation exposure to physicians and improve manipulation stability, but effective guidance remains challenging because conventional fluoroscopic images provide limited spatial information. This paper developed a vascular intervention interactive system that connects a real robotic master controller with the SOFA(Simulation Open Framework Architecture) physical simulation platform and reused successful operation trajectories as visual guidance in physical phantom experiments. Experiments were conducted to validate the master-to-SOFA control loop and to compare three guidance conditions: no guidance, centerline guidance, and successful-trajectory guidance. The control validation results showed reliable command delivery under velocity-mode operation. In the physical phantom experiment, centerline guidance showed no obvious improvement over no guidance. In contrast, successful trajectory guidance provided a clear performance benefit. Compared with the no-guidance condition, it reduced task completion time, collision number, rollback number, and rotation number by approximately 66.8%, 34.3%, 55.6%, and 66.7%, respectively. These results indicate that successful trajectories obtained under instrument-vessel interaction can provide practical guidance cues, offering a simple and effective visual assistance strategy for robotic vascular intervention experiments.
Figures & tables
| Mode | Task | (%) | (ms) | Std. of (ms) | Max. (ms) | |
|---|---|---|---|---|---|---|
| Velocity | Guidewire translation | 300 | 100.00% | 20.25 | 32.40 | 136.70 |
| Velocity | Guidewire rotation | 288 | 100.00% | 15.28 | 31.96 | 173.39 |
| Velocity | Catheter translation | 259 | 100.00% | 27.10 | 38.17 | 155.54 |
| Velocity | Catheter rotation | 270 | 100.00% | 14.50 | 30.36 | 156.75 |
| Velocity | Combined operation | 1106 | 100.00% | 21.64 | 36.94 | 211.55 |
| : number of valid commands; : command-to-pulse success rate. | ||||||
| Condition | (s) | |||
|---|---|---|---|---|
| No guidance | ||||
| Centerline guidance | ||||
| Successful trajectory guidance |
Robotic Endovascular Intervention
Robotic endovascular intervention uses robotic systems to navigate vascular instruments and perform procedures with greater precision, control, or autonomy.
See all 24 papers