A Transparent 3-DoF Assistive Device for Upper-Limb Tremor and Ataxia Compensation: Design and Experimental Validation
Authors: Thomas Dobosz, Veronika Hofmann, Ahmed Barhoma, Ann-Kathrin Henke, Johannes A. Gaus, Hans Mees, Winfried Ilg, Isabell Wochner, Martin Giese, Andreas Bulling, Syn Schmitt, Christophe Maufroy, Urs Schneider, Thomas Bauernhansl, Daniel Haeufle
Organizations: Institute of Industrial Manufacturing and Management, University of Stuttgart, Germany · Biomechatronic Systems, Fraunhofer IPA, Stuttgart, Germany · Hertie Institute for Clinical Brain Research and Centre for Integrative Neuroscience, University of Tubingen, Germany · Center for Bionic Intelligence Tubingen Stuttgart (BITS), Germany · Institute for Visualization and Interactive Systems, University of Stuttgart, Germany · Institute for Modelling and Simulation of Biomechanical Systems, University of Stuttgart, Germany
Assistive devices for patients with neurodegenerative movement disorders must combine high mechanical transparency with the ability to generate corrective interaction forces. Achieving both properties simultaneously remains challenging. This paper presents the completed iAssistADL platform, a 3-DoF assistive device for future tremor and ataxia compensation during activities of daily living. The system employs a transparency-oriented mechanical architecture with all actuators located at the base, resulting in low moving mass, decoupled kinematics, and high mechanical transparency. To compensate for friction-induced transparency limitations along the vertical axis, a hybrid control strategy is proposed that combines open-loop force rendering in the horizontal plane with closed-loop force control in the vertical direction. The platform is experimentally evaluated through transparency measurements and static and dynamic force-rendering experiments at representative configurations. Results demonstrate resistive torques below 0.1 Nm in the horizontal plane, end-effector forces of up to approximately 10 N throughout the workspace, and accurate horizontal force rendering up to 4 Hz. While the vertical axis remains limited by friction-induced dynamics at higher frequencies, the proposed system successfully combines high transparency with controlled force generation, establishing a promising hardware platform for future tremor and ataxia compensation.
Figures & tables
Fig. 1 : User interacting with the iAssistADL platform. The device applies assistive forces to the forearm while preserving natural upper-limb motion.
Fig. 2 : iAssist prototype along with the power supply and real-time computer. Prototype end-effector connected to a 6-D force sensor on a rigid fixture.
Fig. 3 : Schematic of the assistive device with three active DoF. A prismatic joint along the z-axis (blue) actuated via a scissor mechanism, whose motion is constrained by the ball-spline mechanism, a first rotational joint along the ball-spline axis (red), and a second rotational joint bilaterally actuated using cables routed through the scissor mechanism (green). A passive spring mechanism (bottom left) is used to balance the weight of the arm. With joint angles θ1,θ2 , θ3 and motor torques τ1 , τ2 . τ3 . Adopted from [ 18 ]
Force capability
Point
(x,y,z) [cm]
θ1
θ2
θ3
fz,max [N]
Guar. ∥fxy∥ [N]
Max ∥fxy∥ [N]
Anisotropy
detJxy
P1 grasp bottle
(0,50,30)
33.75∘
146.25∘
30.00∘
10.5
9.89
18.2
1.84
0.187
P2 fill glass
(−10,60,25)
52.00∘
146.93∘
24.62∘
10.0
9.89
15.0
1.51
0.202
P3 to mouth
(−10,30,30)
39.00∘
177.86∘
30.00∘
10.5
9.89
28.8
2.91
0.133
TABLE I : Operating points, inverse-kinematic solution, and force-capability analysis. The guaranteed omnidirectional planar force is the shoulder-torque–limited lower bound min(a1,a2) and is configuration-independent : it holds over the entire reachable workspace, not only at P1 – P3 . The pose-dependent quantities are the vertical capability fz,max , the maximum planar force (parallelogram diagonal), the anisotropy (max/guaranteed), and the Jacobian manipulability detJxy .
Fig. 4 : Test positions with glass (G) and bottle (B). F = Front, B = Bottom, L = Left, R = Right.
Metric
Fx
Fy
Fz
Rise time tr (10–90 %)
16 ms
15 ms
3.04 s
Overshoot
33%
33%
28%
Settling time
1.35 s
1.24 s
6.95 s
Steady-state error
−0.04 N
0.08 N
−0.12 N
(relative)
2.1%
5.5%
−16%
TABLE II : Step-response metrics (mean over all cases).
f
Metric
Fx
Fy
Fz
2 Hz
Ar
0.91
0.92
0.98
Δt (ms)
142
155
191
RMSE ϕ (N)
0.36
0.40
1.24
4 Hz
Ar
0.95
0.97
1.87
Δt (ms)
127
124
176
RMSE ϕ (N)
0.68
0.57
3.37
TABLE III : Sinusoidal tracking (mean): amplitude ratio Ar , transport delay Δt , phase-compensated RMSE.
Fig. 5 : Open-loop step response of the Fx axis. Reference (dashed) versus measured force (solid); shaded region is the ±5% settling band.
Fig. 6 : Stick–slip limit cycle on the vertical axis for the Fz=5 N step reference at pose P3. The commanded control torque (top) ramps under integral action while the ball-spline carriage is stuck, breaks away and overshoots on slip, then re-sticks at a new offset, producing a sustained oscillation of the measured force (bottom) about the reference (dashed). The series compliance of the lightweight distal links stores and discharges elastic energy on each slip event, so the integrator hunts across the stiction dead-band without converging to a unique equilibrium.
Fig. 7 : Sinusoidal force tracking on the horizontal Fx axis. Top: the commanded reference (dashed), the raw measured force (grey), and the phase-compensated response (blue). The raw signal reproduces the reference amplitude but is shifted by the near-constant transport delay ( ≈140 ms), which at this frequency manifests as a large phase lag; removing this delay recovers the phase-compensated trace, which closely overlays the reference. Bottom: instantaneous tracking error before (grey) and after (blue) phase compensation, showing that the raw error is dominated by the pure time delay—its removal collapses the error to the residual amplitude- and shape-tracking component (phase-compensated RMSE =0.36 N at 2 Hz), showing that the raw RMSE is dominated by the measured transport delay.
Providing assistance across diverse movements is a central objective of exoskeletons, and anatomical knowledge can enable responsive support that generalizes across tasks. However, anatomical assistance has mainly been studied for lower-limb exoskeletons, where periodic, weight-bearing motions impose lower demands on torque precision. Extending such assistance to complex, nonperiodic upper-limb movements remains challenging. This paper proposes Anatomical Torque with Passivity-Based Control (ATP) for safe upper-limb exoskeleton assistance. First, a scalable musculoskeletal simulation framework trains a unified reinforcement-learning muscle controller that generalizes across upper-limb movements and generates anatomical reference torques without complex biomechanical computations. Second, an online torque-refinement scheme adapts the reference to diverse movements, suppresses tendon-induced spikes, and incorporates a learned anomaly score for safe and comfortable assistance. Third, an interaction torque controller delivers assistance through a cable-driven compliant exoskeleton without constraining motion to predefined trajectories, while an energy tank preserves passivity with theoretical guarantees on torque tracking and system passivity. Simulations and real-world experiments show accurate tracking of long-duration motion sequences and generalization to real-time human movements. The controller achieves accurate torque tracking while preserving passivity and resumes tracking after energy-tank replenishment. An EMG study with five participants further shows reduced target-muscle activity during static and dynamic tasks compared with gravity compensation and open-loop assistance, with reductions of up to 48% relative to movement without the exoskeleton in a dynamic multi-joint task.
Yu Chen, Gong Chen, Xiang Li
Department of Automation, Tsinghua University, China · Shenzhen MileBot Robotics Co., Ltd, China
Contact-rich manipulation requires robots to balance accurate motion tracking with compliant interaction, yet most visual-action policies leave compliance fixed at the controller level. We present Imp-ACT, a methodologically grounded and practical approach to incorporating direction-dependent Cartesian stiffness modulation directly into demonstration collection, without manual stiffness selection or offline target reconstruction. During teleoperation, a self-tuning impedance controller adapts stiffness along the instantaneous direction of motion while maintaining compliance in orthogonal directions. The adapted stiffness is applied and recorded alongside visual observations and motion commands, capturing motion and compliance under the same dynamics. We implement this pipeline using Action Chunking with Transformer (ACT) to predict end-effector pose, gripper action, and motion-direction stiffness from visual, proprioceptive, and wrench observations. The performance of Imp-ACT is evaluated on wiping and plug insertion using both success rate and quantitative measures of contact behavior. Compared with fixed low- and high-stiffness baselines, Imp-ACT achieves comparable or higher success while maintaining low interaction forces. In wiping, it reduces contact-force vibration by approximately 29× relative to the compliant baseline and 180× relative to the stiff baseline. In plug insertion, it reduces forces orthogonal to the insertion direction by 43% relative to the better fixed-stiffness baseline. These results highlight the benefit of maintaining sufficient stiffness along the direction needed for task execution while preserving compliance in other directions to limit contact forces and accommodate environmental constraints.
Luca Zanetti, Doganay Sirintuna, Idil Ozdamar +3
Human-Robot Interfaces and Interaction Laboratory, Istituto Italiano di Tecnologia, Genoa, Italy · Safe AI Lab, Carnegie Mellon University
Safe and efficient assistive planning for visually impaired scenarios remains challenging, since existing methods struggle with multi-objective optimization, generalization, and interpretability. In response, this paper proposes a Momentum-Constrained Hybrid Heuristic Trajectory Optimization Framework (MHHTOF). To balance multiple objectives of comfort and safety, the framework designs a Heuristic Trajectory Sampling Cluster (HTSC) with a Momentum-Constrained Trajectory Optimization (MTO), which suppresses abrupt velocity and acceleration changes. In addition, a novel residual-enhanced deep reinforcement learning (DRL) module refines candidate trajectories, advancing temporal modeling and policy generalization. Finally, a dual-stage cost modeling mechanism (DCMM) is introduced to regulate optimization, where costs in the Frenet space ensure consistency, and reward-driven adaptive weights in the Cartesian space integrate user preferences for interpretability and user-centric decision-making. Experimental results show that the proposed framework converges in nearly half the iterations of baselines and achieves lower and more stable costs. In complex dynamic scenarios, MHHTOF further demonstrates stable velocity and acceleration curves with reduced risk, confirming its advantages in robustness, safety, and efficiency.
Yuting Zeng, Zhiwen Zheng, Jingya Wang +6
School of Information and Software Engineering, University of Electronic Science and Technology of China, Chengdu 610054, Sichuan, P.R. China · Sichuan Provincial Key Laboratory for Human Disease Gene Study, Sichuan Academy of Medical Sciences & Sichuan Provincial People’s Hospital, University of Electronic Science and Technology of China, Chengdu, Sichuan, China. · Research Unit for Blindness Prevention, Chinese Academy of Medical Sciences, Sichuan Academy of Medical Sciences and Sichuan Provincial People’s Hospital, Chengdu, Sichuan, China.