ARTS: An Open-Source, Reproducible Three-Arm Platform for Autonomy Research in Minimally Invasive Surgery
Authors: Jiawei Ge, Chunwoo Kim, Antony Goldenberg, Douglas McDonough, Justin Liao, Bo Ning, Sylvia Ladstaetter, Katya Van Anderlecht, Hao Ding, Laura Connolly, Ethan Kilmer, Francis Fan, Shouyue Hu, Tyler Salvador, Brendan Burkhart, Pit Henrich, Richard Cha, Tito Porras, Mathias Unberath, Kevin Cleary, Axel Krieger
Organizations: Laboratory for Computational Sensing and Robotics (LCSR), Johns Hopkins University, Baltimore, MD 21218, USA · Inner Logic, Baltimore, MD 21211, USA · Korea Institute of Science and Technology (KIST), Seoul 02792, Republic of Korea · Sheikh Zayed Institute for Pediatric Surgical Innovation, Children’s National Hospital, Washington, DC 20010, USA
Research on surgical autonomy needs open platforms that any laboratory can build, yet the most widely shared platform depends on donated clinical hardware, and open alternatives on industrial arms each lack part of the surgical chain. We present the Autonomous Robotic Translational System (ARTS), an open-source three-arm platform for autonomy research in minimally invasive surgery, built entirely from purchasable parts for about USD 137,000. Two UR5e arms hold wristed da Vinci X/Xi instruments, a UR5 arm holds a stereo endoscope with two FLIR cameras, the surgeon console provides haptic handles, footswitches and a stereo viewer, and a hardware stop independent of the software lets a supervisor freeze all three arms in place. The ROS 2 software pins every dependency, installs from the same scripts on bare metal or in containers, and records stereo images with hardware-measured timing. On the real arms, ARTS commands each arm at its controller's full rate (e.g., 500 Hz on the UR5e), the two cameras expose within 12 microseconds of each other, hand motion reaches the arm in 58 ms, and the endoscope image reaches the monitor in 68 ms. The ISO 9283 pose repeatability is 0.14-0.25 mm, comparable to a da Vinci Research Kit (dVRK) Classic arm on the same protocol, and an optical tracker measured a mean remote-center-of-motion error of 0.34-0.37 mm. Three expert and three novice operators performed the Fundamentals of Laparoscopic Surgery (FLS) peg transfer through a laparoscopic trainer box, and one of the experts sutured on a suture pad. The hardware designs, firmware and software are available at https://github.com/MRE-Lab-UMD/arts.
Autonomous robot-assisted surgery demands reliable, high-precision platforms that strictly adhere to the safety and kinematic constraints of minimally invasive procedures. Existing research platforms, primarily based on the da Vinci Research Kit, suffer from cable-driven mechanical limitations that degrade state-space consistency and hinder the downstream training of reliable autonomous policies. We present an open-source, robot-agnostic Remote Center of Motion (RCM) controller based on a closed-form analytical velocity solver that enforces the trocar constraint deterministically without iterative optimization. The controller operates in Cartesian space, enabling any industrial manipulator to function as a surgical robot. We provide implementations for the UR5e and Franka Emika Panda manipulators, and integrate stereoscopic 3D perception. We integrate the robot control into a full-stack ROS-based surgical robotics platform supporting teleoperation, demonstration recording, and deployment of learned policies via a decoupled server-client architecture. We validate the system on a bowel grasping and retraction task across phantom, ex vivo, and in vivo porcine laparoscopic procedures. RCM deviations remain sub-millimeter across all conditions, and trajectory smoothness metrics (SPARC, LDLJ) are comparable to expert demonstrations from the JIGSAWS benchmark recorded on the da Vinci system. These results demonstrate that the platform provides the precision and robustness required for teleoperation, data collection and autonomous policy deployment in realistic surgical scenarios.
Ariel Rodriguez, Lorenzo Mazza, Martin Lelis +4
Department of Translational Surgical Oncology, NCT/UCC Dresden, a partnership between DKFZ, MF UKD, TUD, HZDR Dresden, Germany · Cluster of Excellence-CeTI, TUD, Germany · Department of Visceral, Thoracic and Vascular Surgery, Faculty of Medicine and University Hospital Carl Gustav Carus, TUD, Germany
High-performance surgical master consoles offer intuitive articulated manipulation but are expensive and difficult to reproduce, while accessible commercial haptic devices lack built-in interfaces for surgical wrist articulation and continuous grasp. We present a laparoscopic master in which general-purpose robot manipulators provide the programmable base and surgical-specific interaction is concentrated in a separable distal adapter. Each 6-DoF arm carries a custom 2-DoF direct-drive wrist, continuous grasp sensing, and clutch-based workspace management; the complete bimanual hardware costs approximately US$5,500, below common bimanual commercial haptic references. To support sustained hand-guided use, we characterize current scaling, gravity, and break-away friction and use them for posture-dependent gravity compensation and motion-gated assistance, reducing continuous actuation and thermal loading. A common master state drives either Isaac Sim or an RCM-constrained dual-FR3 platform. Master-state publication to FR3 target update adds at most 8 ms, and tip tracking shows 0.31-0.66 mm mean delay-aligned residuals. In a porcine cholecystectomy experiment, five common operator-side metrics and hand/tip-speed distributions were compared descriptively with human in-vivo telemetry from 120 clinical Versius recordings of 99 cholecystectomy procedures; no interruption originated from the master console, and no abnormal clinical signs were recorded during the 7-day postoperative observation period. These results demonstrate a surgical master built from general-purpose robot arms with surgical-specific interaction concentrated in a separable distal interface.
Minsung Kim, Seonho Shim, Dongho Yee +9
Department of Transdisciplinary Medicine, Seoul National University Hospital, Seoul, Republic of Korea. · Rosota Inc., Seoul, Republic of Korea. · Department of Mechanical Engineering, Seoul National University, Seoul, Republic of Korea. +6
Studying electrosurgical tasks on a research robot requires control of both instrument motion and energy activation, along with a record of when each action occurs. This paper presents an open-source module for da Vinci X/Xi instruments on a UR5/UR5e robotic arm. Five motors transmit torque through timing belts to spring-loaded drive discs. An electrosurgical interposer accepts software commands, gives the surgeon's foot pedals priority, and records the source of each activation request. In bench testing, the interposer recorded all 12 activation and deactivation events from six pedal presses. Ex vivo demonstrations included monopolar and bipolar coagulation of splenic tissue and monopolar cutting during gallbladder dissection. Optical tracking of the Fenestrated Bipolar Forceps showed that calibration brought pitch and yaw gains from 0.904 and 1.14 to 1.00 and 1.02, respectively. The module enables researchers to record instrument motion, video, and the timing and source of electrosurgical commands during laparoscopic tasks.
Douglas McDonough, Justin Liao, Jiawei Ge +6
Laboratory for Computational Sensing and Robotics, Johns Hopkins University, Baltimore, MD 21218, USA. · Department of Surgery, Johns Hopkins Medicine, Baltimore, MD 21287, USA. · Korea Institute of Science and Technology (KIST), Seoul 02792, Republic of Korea.